The Development of Convenient RQ Measuring Device for Patients Real Time Monitoring
YS Kim, HM Jeon, SW Choi, EB Shim
Key Words: RQ, Respiratory Quotient, oxygen and carbon dioxide
Abstract
RQ(Respiratory Quotient) value is obtained from the ratio of the consumed oxygen and the produced carbon dioxide during the patient's respiration. To investigate the efficacy of insulin and diagnosis the metabolic disorder in short time, the RQ value can be used as important parameter. The measurement of oxygen and carbon dioxide amounts is needed large chamber and complex sensors. But If the atmospheric oxygen and carbon dioxide concentrations do not change, the expiratory oxygen and carbon dioxide can be used to obtain RQ value. A convenient RQ measuring device has been developed by using two sensors for O2 and CO2. The estimation of RQ devices confirms that the RQ device can obtain accurate data by eliminating uncertain factor such as delay time and remaining gases.