Yonsei Med J http://www.eymj.org Volume 54 Number 2 March 2013 541

Letter to the Editor http://dx.doi.org/10.3349/ymj.2013.54.2.541

pISSN: 0513-5796, eISSN: 1976-2437 Yonsei Med J 54(2):541-542, 2013

Can We Accurately Assess Liver Fibrosis with Fibroscan

®

Using Fewer Valid Measurements?

Omid Pournik,1 Seyed Moayed Alavian,2 and Saeid Eslami1,3,4

1Medical Informatics Research Center, Department of Medical Informatics, School of Medicine, Mashhad University of Medical Sciences, Mashhad;

2Middle East Liver Disease Center, Tehran; 3Pharmaceutical Research Center, School of Pharmacy, Mashhad University of Medical Sciences, Mashhad, Iran.

4Department of Medical Informatics, Academic Medical Center, University of Amsterdam, Amsterdam, The Netherlands.

Received: December 5, 2012 Accepted: December 6, 2012

Corresponding author: Dr. Seyed Moayed Alavian, Middle East Liver Disease Center, No. 178, Shadab crossroad,

Sepahbod Gharani St. Tehran, Iran.

Tel: 9821-88945186, Fax: 9821-88945188 E-mail: [email protected]

∙ The authors have no financial conflicts of interest.

© Copyright:

Yonsei University College of Medicine 2013 This is an Open Access article distributed under the terms of the Creative Commons Attribution Non- Commercial License (http://creativecommons.org/

licenses/by-nc/3.0) which permits unrestricted non- commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

We read with great interest the article of Jang, et al.1 in March 2012 of Yonsei Medi- cal Journal. The authors challenged the manufacturers’ recommendation for 10 mea- surements for liver evaluation using FibroScan® (Fibroscan device 502-EchoSens, Paris, France). They were seeking the minimum number of measurements which is needed for liver evaluation using FibroScan®. The authors concluded that 3 valid measurements (VMs) would provide the same result as 10 VMs. In our experience in Middle East Liver Disease Center, we agree that in many cases same results might be achieved with fewer numbers of VMs.

There is an important concern in reducing the number of measurements in clin- ic. At present, the detection of reliable evaluations is made based on a success rate of at least 60% and an interquartile range/median value less than 0.3 as reliable.2 Although some researchers believe the above mentioned factors are not predictive of the accuracy,3,4 no other criterion is available. The interquartile range can elimi- nate the effect of outlier and extreme measures, but with only 3 measurements, the results will be affected by such measurements.

Many health conditions can adversely affect the accuracy of Fibroscan®. Steato- sis, metabolic syndrome, high body mass index (BMI), hepatic haemangioma and heart failure are examples of such cases.5 In these subgroup of patients, accuracy of using 3 measurements could be different and lack of subgroup analysis could af- fect generalizability of the results. The average BMI of the participants in this study was in normal range which does not provide much evidence about the diag- nostic accuracy of 3 VMs in patients with high BMI.

If the number of measurements is to be reduced, new criteria for detection of re- liable evaluations should be studied and introduced, especially when there are chances of steatosis, metabolic syndrome or high BMI.

REFERENCES

1. Jang HW, Kim SU, Park JY, Ahn SH, Han KH, Chon CY, et al. How many valid measure- ments are necessary to assess liver fibrosis using FibroScan® in patients with chronic viral hepatitis? An analysis of subjects with at least 10 valid measurements. Yonsei Med J 2012;53:337-45.

2. Lucidarme D, Foucher J, Le Bail B, Vergniol J, Castera L, Duburque C, et al. Factors of ac- curacy of transient elastography (fibroscan) for the diagnosis of liver fibrosis in chronic hepa-

(2)

Omid Pournik, et al.

Yonsei Med J http://www.eymj.org Volume 54 Number 2 March 2013 542

surement by Fibroscan in patients with chronic hepatitis B. Ali- ment Pharmacol Ther 2010;32:498-505.

5. Aalaei-Andabili SH, Mehrnoush L, Salimi S, Shafiei M, Alavian SM. Liver Hemangioma Might Lead to overestimation of Liver Fibrosis by Fibroscan; A Missed Issue in Two Cases. Hepat Mon 2012;12:408-10.

titis C. Hepatology 2009;49:1083-9.

3. Myers RP, Crotty P, Pomier-Layrargues G, Ma M, Urbanski SJ, Elkashab M. Prevalence, risk factors and causes of discordance in fibrosis staging by transient elastography and liver biopsy. Liver Int 2010;30:1471-80.

4. Kim SU, Seo YS, Cheong JY, Kim MY, Kim JK, Um SH, et al.

Factors that affect the diagnostic accuracy of liver fibrosis mea-

참조

관련 문서

Conclusions: WFA-M2BP can support a diagnosis of liver fibrosis with similar diagnostic efficacy to other biomarkers, and predicted liver fibrosis stage ≥2 among patients with

Enhanced liver fibrosis test can predict clinical outcomes in patients with chronic liver disease. Guha IN, Parkes J, Roderick P, Chattopadhyay D, Cross R, Harris S, Kaye P, Burt

This study ex- amined the diagnostic performance of the AsAGP level for advanced liver fibrosis or cirrhosis in patients with chronic hepatitis B (CHB) or nonalcoholic fatty

Patients were divided into groups by the value of liver stiffness measured by transient liver elastography and the cutoff values for severity of fibrosis were derived from the study

Objective: To compare several noninvasive indices of fibrosis in chronic viral hepatitis B, including liver shear-wave velocity (SWV), hyaluronic acid (HA), collagen type IV