981 https://e-kcj.org

A 40-year-old male was admitted to the cardiac care unit (CCU) due to severe pneumonia associated with unstable hemodynamics. Meanwhile, a grade 3–4 systolic murmur was heard at the left lower sternal border. Tracing his history back, he had undergone mitral valve replacement with a Björk-Shiley convexoconcave prosthetic valve (single tilting disc mechanical valve) 25 years before this admission. The reason that he received open heart surgery was infective endocarditis which resulted in mitral valve perforation and systemic embolism. After valve surgery, he continued taking an anticoagulant (warfarin 2.5 mg per day) and was regularly followed up at an outpatient department. Just 6 months before this episode, routine transesophageal echocardiography (TEE) examination still showed the mechanical valve profile was normal and the prosthetic disc worked very well (arrow, Figure 1A). However, the TEE performed at CCU on this admission demonstrated severe mitral regurgitation with absence of the valvular disc acoustic shadow (Figure 1B). Disc dislodgement was therefore highly suspected and emergency surgery was conducted.

During the operation, the disc was found to have escaped from the strut of the Björk-Shiley prosthetic valve (Figure 1C) and was being pumped out of the left ventricle. A new bi-leaflet mechanical valve (St. Jude Medical™ Hemodynamic Plus prosthesis; St. Jude Medical, St.

Paul, MN, USA) was then placed and the patient's hemodynamics were fully restored after the operation.

Dislodgement of a disc occurring in patients with long-term Björk-Shiley prosthetic valve implantation has never been reported. For such a critical event, alert awareness with prompt TEE confirmation and immediate valvular surgery is the only way to ensure patient survival.

Korean Circ J. 2017 Nov;47(6):981 https://doi.org/10.4070/kcj.2017.0125 pISSN 1738-5520·eISSN 1738-5555

Images in

Cardiovascular Medicine

Received: Jun 13, 2017 Accepted: Jul 25, 2017 Correspondence to Ju-Yi Chen, MD, PhD

Division of Cardiology, Department of Internal Medicine, Medical College and Hospital, National Cheng Kung University, 138, Sheng-Li Road, Tainan 70403, Taiwan.

E-mail: [email protected] Copyright © 2017. The Korean Society of Cardiology

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://

creativecommons.org/licenses/by-nc/4.0) which permits unrestricted noncommercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Conflict of Interest

The authors have no financial conflicts of interest.

Author Contributions

Conceptualization: Lin PY, Chen JY; Data curation: Lin PY, Tsai WC, Chen JY; Writing - original draft: Lin PY, Chen JY; Writing - review

& editing: Lin PY, Chen JY.

Pao-Yen Lin, MD, PhD

1

, Wei-Chuan Tsai, MD

2

, and Ju-Yi Chen, MD, PhD

2

1

Department of Surgery, Medical College and Hospital, National Cheng Kung University, Tainan, Taiwan

2

Division of Cardiology, Department of Internal Medicine, Medical College and Hospital, National Cheng Kung University, Tainan, Taiwan

Unusual Disc Dislodgement of a Björk-Shiley Valve after Long-term Implantation

A B C

Figure 1. The transesophageal echocardiographic images (A and B) and surgical removal of mechanical valve (C).

수치

Figure 1. The transesophageal echocardiographic images (A and B) and surgical removal of mechanical valve (C).

참조

관련 문서

[r]

Long-term outcomes of short dental implants supporting single crowns in posterior region: a clinical retrospective study of 5-10 years.. Osteotome sinus

5 ) Stop Light Switch (Mechanical) 6 ) Front/Rear Balance Valve 6 ) Front/Rear Balance Valve 7 ) Pressure Differentiavl Valve 8 ) Brake Warning Lamp.. 9 ) Brake Fluid

5 ) Stop Light Switch (Mechanical) 6 ) Front/Rear Balance Valve 7 ) Pressure Differentiavl Valve 8 ) Brake Warning Lamp.. 9 ) Brake Fluid

Recently, for lumbar disc herniation, Trans-Sacral Epiduroscopic Laser Decompression (SELD) has been a minimally invasive procedure before surgery.. Especially,

Due to flow restriction (lowered volumetric efficiency otherwise) Compression delayed → lowered temperature and pressure rise before combustion. Finite valve times:

The controls shall be water resistant, integrally mounted and shall include a power switch, solid state thermostat, pilot light, vacuum/pressure gauge, safety valve, low

Based on the results of root cause analysis, the following corrective action to mitigate the erosion and corrosion damage at hot leg check valve were