The calibration of x-ray digital tomosynthesis system including the compensation of the image distortion

Young Jun Roha , Kk Won

Koha, Hyung S. Choa, Jin Young J4Jfflb Hyeong Cheol Kimb, Jong-Eun Byunb

aDept of Mechanical Eng., Korea Advanced Institute of Science and Technology bFA Research Institute Production Engineering Center, Samsung Electroncs Co., LTD

ABSTRACT

X-ray laminography and DT(digital tomosynthesis) are promising technologies to form a cross-sectional image

of 3-D objects and can be a good solution for inspecting interior defects of industrial products. It has been known that digital tomosynthesis method has several advantages over laminography method in that it can

overcome the problems such as blurring effect or artifact. The DT system consists of a scanning x-ray tube, an image intensifier as an x-ray image detector, and a CCD camera. To acquire an x-ray image of an arbitrary

plane of objects, a set of images (8 images or more) should be synthesized by averaging or minimally calculating point by point. The images, however are distorted according to the configurations of the image intensifier and the x-ray source position. To get a clear and accurate synthesized image, the corresponding

points in the distorted images should be accurately determined, and therefore, precise calibration of the DT system is needed to map the corresponding points correctly. In this work, a series of calibration methods for the

DT system are presented including the correction of the center offset between the x-ray and the image intensifier, the x-ray steering calibration, and the correction of the distortion of the image. The calibration

models are implemented to the DT system and the experiment results are presented and discussed in detail.

Keywords : x-ray laminography, cross-sectional image, digital tomosynthesis, x-ray calibration

1. INTRODUCTION

X-ray technology is a good solution for inspecting inner defects of industrial products. In case of conventional x-ray radiography images, however, it is very difficult to correctly recognize the internal shape of objects, because all objects on the projection line are overlapped in the images. On the other hand, x-ray cross-sectional imaging systems forming an arbitrary cross-section image of a 3D object make it easy to inspect the inner shape or structure of it. PCB solder joint inspection is one of the industrial fields which need such an x-ray inspection system. Recently, new SMD(Surface Mount Device) technologies are developed and new electrical packages are made utilizing such as BGA(Ball Grid Array), FCA(Flip Chip Array) and J-type lead. One drawback of these packages is that it is unable to inspect those chips by visual inspection system, because the solder joints are hidden under the chips themselves. Furthermore, it is difficult to inspect them by conventional x-ray transmission image to solve the overlapping problem. Therefore, the x-ray cross-sectional

imaging system promises to be a good solution for these difficult problems.[1]

An x-ray cross-sectional image is integrated from a number of images projected from different directions. Several

Further author information -

H.S.Cho: E-mail : hscho@lca.kaist.ac.kr; WWW: http://1ca.kaist.ac.kr/-hscho; Telephone +82-42-869-32 13; Fax : +82-42-869-32 10

PartoftheSPIE Conference on Machine Vision Systems for Insoection and Metrolonv VII

248 Boston.Massachusetts • November 1998 SPIE Vol. 3521 • 0277-786X/98/$1 0.00

(2)

methods such as tomography, laminography and digital tomosynthesis have been used to achieve this. Tomography has been mainly used in medical area, but is being applied in industrial fields such as precision inspection of casting products in these days.[2]

Laminography is another method for acquiring cross-sectional images, and is originated by Bocage[3]. Its principle comes from the geometric focusing effect by the synchronized motion between an x-ray source and a detector, and are showed in the Figure 1.

A

(A+B+C+D)/4 B

Figure1 .Theprinciple of Laminography or DT

Laminography is applied to inspect PCB solder joints by Adams[fl, Black[4], Rooks[5,6]. The basic principle of digital tomosynthesis is the same as that oflaminography, but the difference is in integrating images projected from different views by software not by hardware[7]. But in general, the quality of the cross-sectional images is not as good as that of the radiography images in spite of its advantage. It comes from the fact that it is a synthesized image from a number of images.

To improve the quality of the x-ray images, precise calibration of the system including the compensation of the image distortion and image integrating methods are needed.

In this paper, three major topics are presented. First, the newly developed DT system will be introduced. Secondly, a series of calibration methods for the system are presented. Finally, x-ray cross-sectional images of PCB solder joints acquired from this proposed system are demonstrated, and analyzed in detail.

2. SYSTEM CONFIGURATION

Figure 2 shows the structure of the DT system developed, which is composed of a scanning x-ray tube, a image intensifier, a rotating prism and a zoom camera. To project x-ray to an object(PCB) at different directions electrically, a scanning x-ray tube is used instead of a rotating x-ray tube mechanically. And an image intensifier with a large input screen is used as a x-ray detector so as to get all images projected at various directions. The region of interest of the PCB is projected on a circular trajectory on the image intensifier as the x-ray is steered on a circular trajectory, and eight or more images are sequentially acquired by the zoom camera through the rotating prism. To catch the projected images on the trajectory, the prism rotates to synchronize with the x-ray motion. The captured images are saved in the digital memory and synthesized after 1 rotating of the x-ray and the prism [8].

The x-ray imaging conditions in the system such as the projection angle and the magnification are determined by the geometric relations of the parameters which are the radius of the rotating x-ray and the distances from x-ray to the object plane and to the surface of the image intensifier. The projection angle is an important parameter in the DT system, since the artifacts that cause distortions of the cross-section in the resulted image can be reduced or removed as the projection angle increases in general [9,10,11].

D

X-ray B

Objects

B

C X-rayDetector

(3)

Soanng X-y tuDe

POB

XYZ

_____

//

mage

1'

Zoo' camea

> Ro p'-s

Figure 2. Digital tomosynthesis system

The rotating prism shown in Figure 3 makes it possible for the zoom camera located at the center to get the sequentia' images on the circular trajectory. By configuring the system in this way, we can get x-ray projection image with different directions or various projection angles and magnifications. The principle is that light is refracted and shifted its path as it pass through another matter. If we put a prism in front of the camera and rotate it around the camera center axis (Figure 3.

(b)),the images which is located on the circular path with a radius are acquired through the camera.

(a) Light path shift by refraction (b) Local images acquisition by a rotating prism Figure 3. Principle and configuration of Rotating prism

Image ntcsc

— ."t'

L

(4)

3. X-RAY

SOURCE CALIBRATION 3.1 X-ray center alignment

In the DT system developed, an x-ray source is steered on the circular trajectory with its rotating axis centered at the center point of the inspection region of interest in the PCB. For the symmetric and consistent imaging condition for 8 images projected on the image intensifier, the center of the rotating x-ray source needs to be coincident with the center of the image intensifier. To achieve this we need to defme a reference coordinates which can locate all the points of the x-ray tube and image intensifier. For convenience, let us assume that the reference coordinate is chosen to be that of the image intensifier ; the origin is defmed by the center point on the input surface of the image intensifier and the three axis coordinates are chosen as shown in the Figure 4. In general, there exists an offset between the center line of the x-ray source (a point-dashed line in the Figure 5)andthat of the image mtensifier(a solid line in the Figure 5).

Tofind this offset value, it is assumed that the x-ray is a point light source. In case of a point light source, the objects on the off-axis (the points a and c on the planes 1,2 in the Figure 5)areprojected onto different positions as the distance from the light source to the objects are different except the point that is on the center line (the point b in the Figure 5).

1

surface

of 2

intensifier

Image plane

Figure 4. Center offset of the system Figure 5.Thecharacteristic ofthe point light source

Figure 6 shows the procedure to fmd the offset value by updating x-ray position in such a way that x-ray approaches to the center axis of the image intensifier. A 2mm diameter sized steel ball 'A' is fixed at the center of the input surface of the image intensifier and another 0.7mm diameter sized steel ball 'B' is positioned on the x-y table. When we project x-ray to the two balls, we can acquire an x-ray image which includes two shadowed points by the ball A and ball B. And by using the projected positions of the two balls, the offset value can be found.

At the first step, shown in Figure 6 (a), the x-y table is controlled to move the point B in the direction of ball Aposition in order to make the point B coincide with the point A in the x-ray image. Then, the x-y table is elevated 50mm upwards as shown in Fig 6(b), and the positions of two balls in the image are checked. If the center axis of the x-ray source is not located on the that of the image intensifier, as shown in the Figure 6 (b), the point B' which is the projected point of Bis separated from the point A in the image. In case of that, the x-ray source is steered from d to d+1 in order to locate the two points at an identical position. And the point B shown in the Figure 6 (c) is lowered to the plane 1 again and thesame procedures (a)—{d) are carried out repeatedly. To fix two points A and B in the x-ray image, P1 feedback controllers are used in the step 1, x-y stage motion control, and in the step 2, x-ray steering control. This iterative method makes the x-ray

source approach to the center axis eventually and the control value(steering voltage) here is to be the offset value for x-ray source to be located at that of the intensifier.

x-ray tube

point light source

offset

(5)

(a) step 1

Figure 6. X-ray center alignment procedure

Figure 7 shows the variation of the x-ray source steering coordinates when the x-ray is steered to the origin by a number of iterations. The experiment is performed under the condition that the tube voltage and the current are 1 10kV and O.O3mA, respectively. The result show that the adjustment is terminated at the 1 1thiteration.In the figure 7(a), C indicates the x-ray steering coordinate at the thiteration;e.g. C' the first iteration and 'the1 1thiterationwhich is the fmal one in this case.

This situation is well explained in Figure 7(c). As can be seen from the diagram, regardless of the location of the planes parallel to the image plane, the point B on the center axis is always projected unto the identical point A.

-0.75 . . .

— —

.080 -••- 1110

11 :•—4.' : :

-O.85 .

8 ... ,..

. .

090

\6

.0.95 . . . . ..

. .

-1oo. ,

- .

:

NNNN.3

: Ni

—14C I I I I I I

-2.55 -2.50 -2.45 -2.40 -2.35 .2313 -225 -220 -2.15 -2.10

x-axis steering coordinate

(a) Experimental result (b) 1st iteration (c) 1 1th iteration

3.2 X-ray steering calibration

Figure 7. An experimental result ofthe x-ray center alignment

252

The x-ray source is steered by 13 bit digitized signals of DC voltage(-1O +10 volt) for each x, y axes. Since the principle of the x-ray source steering is that the control signals deflect the electron beam insidethe x-ray tube, the amount of the deflection is related to the kinetic energy of the electron and varies with the power applied to the tube. This means that the steering signals are dependent on the x-ray power condition and thus need to be calibrated to the real deflection(or

steering) value. The x-ray steering position is hard to detect directly, but can be analyzed indirectly from the x-ray image. In Figure 8, L1 and L0 are the distances from the x-ray source to the image plane and to the object plane, respectively,and from

center axisi ' x-ray axis x ray tube 1 C

:

B planel : x-Ycontrol

I

i I C

B1

plane2

z-motion

ft:

114 .

'! :

I I

Ct+1 j : X-ray steering

Ctlj I

B!

: z-motion !

L

. Bj : planel

:

A

Imageplane &__

A B---5,-

A A

(b) step 2 (c) step 3 (d) step 4

. 0

C0

.

lID

—plane 2

C" 1

B'

B

A

plane 1

Image plane

(6)

an arbitrary the point Q(Xq ,Yq) we can calculate the x-ray position P(x ,

y )

by its geometric relation, which is represented by the equation (1) given below. The aim ofthis calibration is to fmd the relationship between the x-ray position and the control signals applied to the x-ray tube. After a steel ball on the object plane is moved to a known position

Q(Xq ,Yq) the x-ray source is steered and positioned to locate the projection point by the ball at the center point 0 of the image intensifier, as shown in Figure 8. Then, the control signals at that time can be related to the x-ray positions calculated from the equation (1). Figure 9 is a result of the calibration under the power condition of I 10kV and O.O3mA. The experiment is performed for 7 different positions of the x-ray source, and the x-ray control signals are linearly related to the x-ray source positions. The resulting scale factor is 0.256 and this means that x-ray is steered by 10mm with the control

signal 2.56 volt.

.

,r'

'I

p

, ' x= 0.256;

7

.

..

, .

. .

.

-20 -15 -10 -5 ) 5 10 15 20

X-ray

source position( x )' mm

Figure 9. A result ofthe steering calibration

[xpl L -L0 [xq

(1)

[YP] L [Yqj

4. IMAGE DISTORTION

In the DT system, the region of interest on the PCB is projected unto the peripheral area on the input surface ofthe image intensifier as the x-ray source rotates around the center axis. The x-ray images are distorted both in intensity and shape by the curved input surface of the image intensifier as shown in Figure 10. The distortions in the images are affected by the direction of the projected x-ray. Figure 1 1 shows the eight distorted images of the uniformly spaced grid on the image intensifier as the x-ray projects it at eight different directions.

In acquiring a cross-sectional image by synthesizing these eight images, a problem arises that the correspondence of each point of the images are not clear, since the raw x-ray images taken from the image intensifier are distorted. Thus, correction of the distorted images are preceded before synthesizing them[12]. The first distortion issue is on the intensity distortion.

Intensity distortion means that the intensities are not uniform in the x-ray image as shown in Figure 12 (a), and it is caused by the different x-ray incident angle onto the curved surface of the image intensifier. The second distortion issue is on the shape distortion. The curved input surface of the image intensifier distort not only the intensities but also the shape of the

images projected on the surface; the peripheral area appears expanded in comparison with the central area as shown in Figure 1 1 .

If

the symmetric and consistent imaging conditions for the eight images are guaranteed, the distortion can be modeled and corrected from one sampled image ofthem, Then the image that is acquired at the direction of 90° can be used for this corrective work. To overcome the distortions of the x-ray images, we propose a series of correcting methods and their correcting procedures. The raw image is corrected for intensity distortion before it is corrected for shape distortion as

shown in the Figure 12.

X-ray

tuberfPYp)

-2

Figure 8. The x-ray steering calibration

(7)

'-ri

Figure 10. Image distortion by the image intensifier Figure 11. Eight distorted images obtained from different views

A

(a) distorted raw image (b) intensity corrected image (c) shape corrected image

4.1 Intensity distortion compensation

Figure 12. Image distortion compensation

254

The central area of the image intensifier is brighter than the peripheral area, since the incident angle of the x-ray is getting declined toward the peripheral area of the image intensifier. In the off-axis images, the point nearest the center of the image intensifier has the maximum intensity and the intensity of the point toward the peripheral area is getting degraded.

To compensate the distorted intensities and make them uniform in the image, the degraded intensities are regulated at the level of the maximum intensity in the image.

The distorted intensity Id

(i, J)

at

a point (i, J)

inthe image can be corrected to it's compensated intensity 1)(

(i, j)

bydividing the compensating ratio function f(Ld) forthe point, as given in the equation (2). lIere L, (i, j)isdefined as the distance in the image from the highlight point H (i,1 j11) tothe point (i, J)considered for compensating the intensity.

The ratio function f(Ld) hasa value decreasing with Lj and has a value between 0 and I, thus plays a role of modifying

an intensity id

toan increased values of cI)e.

A

4.,..,,. * MSit*•..

Intensity distortion compensation

Shape distortion compensation

(8)

f(Ld )

f(Ld) = C0 + c1 L1 + c.

+

Ld(i,j)

=

+(j

)2

(2)

(3) (4)

To build the ratio function f(L1 ), 10 different values of intensity are sampled from the 10 small areas on the path from the point H to the point L. as shown in the Figure 13. A polynomial model is applied to describe the ratio function

f(Ld

) accordingto the distance Ld in the image, which is represented by the equation (3). 'l'he coefficients ot the polynomial arc determined by a least square method.

I)ark area

41X1 0(X) 6(9)

The shape distortion can be corrected by finding a mapping relationship between the uniformly spacing grid image and its distorted image. Thus, a point (i,j) in the original undistorted image is mapped to a point (x. in the distorted image by the mapping relation. In this research, a 2 dimensional mapping function is built by using the sampleddata sets which consist of the uniformly spaced grid points with II rows and 11 columns and their corresponding points in the distorted images. Figure 14 shows the relations between the points of the original grid (a) and the points of the distorted image(h).

There are two main features in the distortion and a simplified distort model can be made based on these characteristics. One is that the distorted grids are always symmetric with the projection center line of the x-ray in the images.The other is that the points on the peripheral area are more elongated than the points near the central area of the image intensifier.

Based on the two features in the distortion, five polynomial equations, from the equation (5) through (9) are proposed to represent the relation between the two corresponding points in each image. A horizontal line inthe original image, which is an orthogonal line with respect to the projection center line, is distorted and become a curved line. It canhe modeled as a 2' order polynomial function in equation (5)with

a peak point (x,P)

onthe projection center line and a coefficient (1 as shown in the Figure 14. Therefore, i coordinates of the points in the original image determine the two variables P and (I which represent a curved line in the distorted image. To generalize the P and a as functionof i, the data sets which consist of the 11 rows E6 (k1.2 11) in the original image and the same number of values for P and are acquired and titted by 3rdorderpolynomial functions of i, defined by equations (7)and(8).

H Bright area Dark

area

4.2 Shape distortion compensation

f(L1)

N

10 09 08

max 01

06 05 04

Figure13. Distorted intensity profile Bright area

0 1W 2W 3W

Ld

(9)

150 200 250 300 350 400 450

[TI

x

4

y4

(a) Original grid points (b) Distorted grid points

256

Figure 14. Two-dimensional point mapping

y(i, j) = a(i) {x(i, j)

— x 2

+ P(i) x(i, j)

= /3(1)

j + j

P(i)=p0 +p1 .i+p2 .i2 +p3 .i3 a(i)=a0 +a1 •i+a2 •i2 +a3 •i3 /3(i)=b0 +b1 •i+b2 2

+b3

.3

(5) (6)

(7) (8) (9)

And the jcoordinatevalues of the points on a horizontal line in the original image are linearly related to the x coordinate values of the points on its distorted curved line, and the relation is represented by equation (6). The coefficient fi of the equation is also a function of i, since the curved line is determined by the coordinate I in the original image. p is also defmed by a 3rdorderfunction of 1 as a given in equation (9), and its coefficients b0 -1,, are determined by a least square fit of the data set which consists of the 1 1 rows k the original image and the same number of f3k acquired from the distorted lines.

The distortion model is completed by fmding the constant coefficients a0 a3 ,b0

-

b3 and p0 -p3 , andas the result of the model, an arbitrary point (i, j) the original image can be mapped to the points(x, y) in the distorted image by the equations (5) and (6), where the coefficients ofthe functions P(i) ,a(i) and fl(i) are pre-determined by the coordinate 1.

5. X-ray images

In this section, several DT images of PCB are presented to show validity of the cross-sectional imaging method and confirm accuracy of correcting the distortion of the images. The DT images are acquired by integrating the 8 images taken from 8 different off-axis images which are corrected by the distortion-correcting method explained previously. The experimental conditions are represented by Table 1, which are the same conditions used in the calibrations.

Projection center line Projection center line

—.

'N

- •-•

Jc 50

100 150 200 250 300 350

.

lx 3/ a1(x-x5)2+P,

••.

• • J

•.. • . .

S S •4 • S• •

•... .11....'

.I• • • 41 • • S I

— ..'

&_,_1_._q_*.._.-.-Y axx)2+Pk

. E;iii y a(xx2+PN

50 100 150 200 250 300 350 400 450 500 550 600 0 50 100 150 200 250' 400 450 500 550 600

400 450

(10)

Parameters Values

X-ray power 110kv. 0.O3mA

Projection Angle 28°

Magnitude

-__________

7

Field of view 7mm x 7mm

Number of Images integrated __________________________

Table 1. The experimental(imagmg) conditions

Figure 15 (a) shows a conventional x-ray radiography image of a double sided PCB with rectangular chips Ofl both sides of the board. In the radiography image, the chips that are attached on the two surfaces appear overlapped and hard to discriminate. On the other hand, Figures 15 (b) and (c) show DI images for the top and the bottom surface of the hoard respectively, and 3D shapes of the solder joints on each side of the PCB can be reconstructed from the l)T image, as shown

in Figure 15 (d). These DT images provide a good condition for inspectmg solder joint quality.

(a) Radiography image (b) DT image of the top surface (c) DT image of the bottom surface (d)Reconstructed 3D shapes of the solder joints

Figure 5. x-ray images for double sided PCB

Figure 16 shows the x-ray DI images of BGA(Ball Grid Array) with lead balls under the chip package. In the slice images. the dark regions represent the cross-section of the focal plane, and the shadowed regions near the dark areas ate the artifact affected by the materials located out of the focal plane. As the focal plane is moved another slices of it, the cross- section of the lead ball or the solder joints appear differently; the cross-section at the middle of the lead ball has the maximum diameter and the diameters at the other focal planes are smaller than that.

0000 0®G0

8 images

. 'p.. A

A

a as S

A

B

'

Ohp Package

Focai plane

E— Z1;QC Q_ . .

Lead /

BaH solder O;ntS

H

H H

13(.\ !ai (rid .\r1\ Ib) I )i iniugcs lo the oCci plane

- 3

(11)

258

These cross-sectional images provides a good que or unique solution to inspect the objects whose shapes are very difficult to obtain through imaging techniques, such as BGA or J-lead type chips. The artifact in the image may bring about some confusion to determine the boundary of the cross-section. But, the artifact can be reduced by some imaging conditions in the DT system, such as x-ray projection angle, the number of images to be integrated and image synthesizing method.

Those factors are discussed in detail in the references [9,11,14-16].

6. CONCLUSIONS

In this work, an x-ray DT (Digital Tomosynthesis) system for solder joint inspection has been developed, and a series of calibration works for the system have been presented. By the DT system 8 off-axis images are sequentially taken as the motion of the prism located under the image intensifier synchronizes with the rotating motion of the x-ray. They are integrated to be a slice image of an object.

In generating a cross-sectional image by the system, there are two major problems to be solved. One is how to steer the x-ray to the desired positions on the predetermined circular trajectory, the center of which is coincident with that of the image intensifier. To realize it, alignment of the two center axes of coordinates, the x-ray and the image intensifier, has to be preceded before we calibrate the x-ray steering position in proportion to the control voltages. To align two coordinates, an iterative method to drive x-ray to be positioned on the center axis of the image intensifier is used. Based on the integrated coordinate system, x-ray steered positions are calculated indirectly by use of the geometric relations between the positions of the object and its projected position on the imaging surface. The other problem is how to correct the distorted images due to the curved input surface of the image intensifier. To correct the distorted x-ray image in intensity and shape as well, a simplified distortion model was built by use of the uniformly spaced grid and its distorted image. The distorted images are corrected by the distortion model and the results show that the corrected images are nearly the same as the original ones within 2 pixels error in the image of 640x480 pixel.

The resulted x-ray DT images show the usefulness of the cross-sectional imaging technique for PCB solder joint inspection. In the example of double sided PCB, the slice images for the top or the bottom surface of the PCB are not affected by the chips on the opposite surface, and thus the defects of the solder joints can be easily found. In the application of BGA chip inspection, the cross-section images of the BGA chips acquired by the DT system is a unique visual information for the inspection. Based upon these fmdings the proposed method may be used as an useful information to judge quality of the solder joints.

ACKNOWLEDGEMENTS

This work was supported by Samsung Electronics Corporation and conducted through 1996-4997. The authors would like to thank Drs. H.N. Joo, and S.K. Kim of Samsung Electronics Company for their valuable advice.

REFERENCES

I . Adams,"X-ray laminography analysis of ultra fme pitch solder connections on ultra-thin boards", Integrated Circuit Metrology,Inspection, and Process Control V (SPIE) Vol.1464. 1991, pp 484-497

2. "Computed tomography details casting defects", Advanced Materials & Process, Nov. 1990, pp 54-56.

3. E. M. Bocage, French Patent 536464(1922)

4. Black, D. L. Millard, and K.Nilson, "An animated interface for x-ray laminographic inspection of fme-pitch interconnect", IEMT Symposium 1991 ,pp207-211

5. Rooks,M. Okimura and R. Urban, "Inspectionofvery-fme pitch connections on PCMCIA cards", Proceedings Nepcon West'93, pp 752-762, 1993.

6. M.Rooks, B. Benhabib, and K. C. Smith, "Development of an inspection process for ball-grid-array technology using scanned beam x-ray laminography", IEEE trans. on Components, Packing, and Manufacturing Technology -PartA.

Vol. 18, No 4, December 1995. pp 85 1-861.

7. Rooks and T. Sack, "x-ray inspection of flip chip attach using digital tomosynthesis", Surface Mount international conference (San Jose) 5 1-55,1994

8. Y.J.Roh, The analysis and design of x-ray cross-sectional imaging system using digital tomosynthesis, M/S thesis, Korea Advanced Institute of Science and Technology, 1997.

(12)

9. Y.J.Roh, S.T.Kang, H.S.Cho, H.C.Kim, S.K.KiIII, "Analysis of x-ray image qualities-accuracy of shape and clearness of image —usingx-ray digital tomosynthesis", '97 Korean Automation and Control Conference, Nov. 1997, pp. 719- 723.

I 0. T. A. Siewert and Mark W. Austin, "The contributions of out-of-plane materials to a scanned-beam laminography image" Materials Evaluation, Oct. 1994, pp 1194-1198.

I 1 . S.T.Kang,Arbitrary cross-sectional recognition of 3 dimensional objects using a digital tomosynthesis, Ph.D. thesis, Korea Advanced Institute of Science and Technology, 1998.

1 2. Development of an PCB solder joints inspection system using x-ray, research report, Samsung electronics, 1997.

13. S.T. Kang and H.S. Cho, "The Quality Evalution and Improvement for Digital Tomosynthesis Images", Materials Evaluation, 1998.

14. S. T. Kang, H. S. Cho, "A projection method for reconstructing X-ray images of arbitrary cross-section", NDT & E International, 1998 (to appear).

15. S.T. Kang, Y.U. Kim and H.S. Cho, "A digital tomosynthesis method for evaluating the soldering states of ball-grid- array joints", 7th AnnualResearch Symposium of Transfer of Emerging NDE Tdchnologies, 1998.

16. ST. Kang, J.H. Jeong, H.O. Song and H.S. Cho, "A new X-Ray cross-sectional imaging system for arbitrary angle inspection of BGA package", Proc. of NEPCON EAST '97, pp.109-120, 1997.

참조

관련 문서

Modern Physics for Scientists and Engineers International Edition,

한편 이 연구의 궁극적인 목적인 서비스 가격의 적절성을 평가하기 위해 질병 의 중증도를 통제한 상태에서 소득수준에 따른 외래이용을 분석하기 위하여

Five days later, on 15 January 1975, the Portuguese government signed an agreement with the MPLA, FNLA and UNITA providing for Angola to receive its independence on 11

Usefulness of co-treatment with immunomodulators in patients with inflammatory bowel disease treated with scheduled infliximab maintenance therapy.. Oussalah A, Chevaux JB, Fay

Inclusion and Inclusiveness: Shared Vision of Youth for Local, National, and Global Village Inclusion at large stands for embracing populations with disabilities and

그것은 이미지를 문자적 으로 서술하는 문제이며 옐름슬레브Hjelmslev[4]의 용어를 쓰자면 (코노테 이션과 반대되는 의미로서) 공정작용operation의

웹 표준을 지원하는 플랫폼에서 큰 수정없이 실행 가능함 패키징을 통해 다양한 기기를 위한 앱을 작성할 수 있음 네이티브 앱과

_____ culture appears to be attractive (도시의) to the