Evaluation of skin surface dose for head and neck cancer patients treated with intensity-modulated radiation therapy using in vivo dosimetry

N/A
N/A
Protected

Academic year: 2021

Share "Evaluation of skin surface dose for head and neck cancer patients treated with intensity-modulated radiation therapy using in vivo dosimetry"

Copied!
2
0
0

로드 중.... (전체 텍스트 보기)

전체 글

(1)

2011년도 춘계학술발표회 논문요약집 대한방사선방어학회

84_http://www.karp.or.kr

Evaluation of skin surface dose for head and neck cancer patients treated with intensity-modulated radiation therapy

using in vivo dosimetry

김 연 실 ․ 이 동 수 ․ 강 영 남 ․ 유 미 나 ․ 홍 주 영 ․ 윤 세 철 ․ 장 홍 석

가톨릭대학교 서울성모병원 방사선종양학과 E-mail: yeonkim7@catholic.ac.kr 중심어 : 피부선량, in vivo dosimetry, 두경부암, IMRT

서 론

Use of intensity-modulated radiation therapy (IMRT) for head and neck cancer is gradually increasing, because it could facilitate more sophsticated treatment of target volumes and reduction of acute and late sequelae. However, theoretically, there is a potential risk of increased skin surface dose resulting from multiple obliquity effects caused by multiple tangential beams.

Moreover, we sometimes confronted with more skin reactions in the patients treated with IMRT than conventional techniques. In this study, we evaluated skin surface dose adjacent to the target volumes to verify whether the use of IMRT would increase the skin dose more than we predicted.

재료 및 방법

A total of three patients who were treated with IMRT for head and neck cancer was

included in this study. Disease sites were as follows : one with base of tongue cancer, one with tonsillar cancer, and one with nasopharyngeal cancer. Median age was 56 years. All of the patients were treated to the cervical lymph node regions due to the gross nodal disease and/or elective nodal irradiation.

Treatment planning was performed by nine field IMRT for all of the three patients. We measured skin surface dose at the different three points per each patient with maximum two times. In vivo dosimetry was performed with MOSFET.

결과 및 고찰

Prescribed target dose of each patient was 160 cGy, 212 cGy, and 210cGy, respectively.

Mean measured surface skin dose of each patient was 39.83 (range, 23.9-64.1) cGy, 59.7 (range, 51-66.5) cGy, and 49.07 (range, 41.5-61.7) cGy, respectively. Mean measured surface skin dose was mean 25.5 (range, 14.9-40.1) percent of prescribed adjacent target dose.

(2)

2011년도 춘계학술발표회 논문요약집 대한방사선방어학회

제4분과(의료 및 생물)_ 85

결 론

This study had shown that the use of IMRT did not increase the skin surface hot point dose.

The measured skin surface dose was 20 to 40percent of the adjacent target prescription dose, and was within acceptable dose range. Our study had some limitations with small number of experimental patients and methodological problems. Potential risk of increasing skin dose with bolus effect of aquaplaster should be examined in the future trials. In addition, the accurate set-up verification should be maintained because of steep dose gradient between skin surface and target volumes within a short distance in the head and neck cancer patients.

참 고 문 헌

1. E.J. Hall, Radiobiology for the Radiologist, 5th Ed., pp. 124-135, Lippincott, Philadelphia (2000) 2. 2. Qi ZY, Deng XW, Huang SM, Zhang L, He ZC, Li XA, Kwan I, Lerch M, Cutajar D, Metcalfe P, Rosenfeld A. In vivo verification of superficial dose for head and neck treatments using intensity-modulated techniques. Med Phys. 2009 Jan;36(1):59-70.

3. Varadhan R, Miller J, Garrity B, Weber M. In vivo prostate IMRT dosimetry with MOSFET detectors using brass buildup caps. J Appl Clin Med Phys. 2006 Nov 28;7(4):22-32.

참조

관련 문서

Patients with breast cancer; ovarian cancer; renal cell carcinoma; pancreatic neuroendocrine cancer; colorectal cancer; head and neck cancer; non-small cell lung

Select: Skin surface (Target of intersection, white border on upper figure) Click: [GO] (in the lower right side of modeling window). Select: original and duplicated curves

Micro- and nano-sized pores were formed on the surface of the alloy using PEO and anodization methods, and the pore shape change according to the Zr

• Concentration gradient of the oxidizing component, H 2 O 2 in the etch solution close to the surface and inside the thin surface oxide, during a wet chemical

• In-situ gamma-ray spectrometry for radioactivity analysis of soil using NaI(Tl) and LaBr 3 (Ce) detectors/ Radiation Protection Dosimetry, 2019. • In-situ CeBr

In this thesis, experimentally evaluated the characteristics of MOSFET dosimeter for reproducibility, linearity, isotropy, dose rate dependence and verification

Therefore, nurses should be responsible for the relations to resolve the needs of care between nurses and patients and through communication with the patients,

In the analysis of the dose constraint of nuclear medical radiation workers using chi-square test statistics, the dose constraint at (the time of) the highest