Laboratory Establishment for Measuring Uranium Radioactivity in Urine

Cheol Hyun Kang* and Jong Il Lee

Korea Atomic Energy Research Institute, 111, Daedeok-daero 989beon-gil, Yuseong-gu, Daejeon, Republic of Korea *

[email protected]

1. Introduction

Recently .RUHDQJRYHUQPHQW¶VUHJXODWRU\ guideline suggests bioassay of radiation workers who can have annual dose of over 1 mSv [1]. When it comes to radiation workers who deal with Uranium, it is impossible to achieve minimum detectable activity through in-vivo bioassay, like lung counter. In-Vitro bioassay can be a solution for achieving lower MDA. Nowadays, many research groups have been developed a variety of in-vitro bioassay methods for Uranium bioassay. For many of those use the 1-day urine of the radiation workers. However, it is hard to gather 1-day urine of the workers due to cumbersome reasons. The creatinine or volumetric normalization can be alternatives even though there are variation from individual or different urine samples [2]. The urine sample should be acid digested, because organic matters can interrupt measuring the uranium concentration by producing complex compounds. After eliminating organic matters, the solid phase extraction (SPE) process can be executed. Mostly, UTEVAŠ Resin, Eichrom is used for extracting Uranium. The extracted uranium can be measured either alpha spectroscopy or Ion-Coupled Plasma mass spectroscopy (ICP-MS).

In this study, bioassay laboratory to access the internal dose of workers who deal with Uranium isotopes has been established. The laboratory will FDUU\ RXW ELRDVVD\ RI .$(5,¶V UDGLDWLRQ ZRUNHUV who treat Uranium isotopes.

2. Analytical Methods

2.1 Microwave Digestion System

7R UHPRYH RUJDQLF PDWWHUV .$(5,¶V ELRDVVD\ laboratory adopted ETHOS EASY Advanced Microwave Digestion system, Milestone srl. It can make high pressure and temperature condition so that the urine samples can be digested faster than hot-plate assisted wet digestion without the loss of volatile materials in the urine samples. The urine samples were kept in the 3ଇ refrigerator and 5 g spot urine sample is accurately weighed. The sample is transfer to the PTFE digestion vessel to digest with 8 ml concentrated HNO3. The sample put into the Microwave system and then digested under the setting program (Table. 1) under the optimal conditions, the clear digestion solution can be obtained. The digestion solution transferred to the glass beaker and evaporation to dryness in 140ଇ hot- plate. The residue is dissolved with 2 mL 4M HNO3 for the solid phase extraction (SPE) process.

Table 1. Microwave digestion program for urine samples Step 1 Step 2 Step3 T/ଇ R time /min H time /min T/ଇ R time /min T/ଇ R time /min H time /min 120 5 3 170 5 200 7 15

Note : R means increasing temperature time H means holding temperature time

(2)

354

2019 ⦽ǎႊᔍᖒ⠱ʑྜྷ⦺⫭⇹ĥ⦺ᚁݡ⫭םྙ᫵᧞Ḳ

2.2 Solid Phase Extraction

The one-stage ion exchange column chemistry procedure has been used in order to remove interfering elements except Uranium isotopes. The SPE begins with 2 mL empty column (Eichrom), which were packed with 1 mL of UTEVAResin (Eichrom). The 2 mL of the sample solution is loaded onto the ion exchange column. The complete ion exchange chromatography procedure is presented in Table 2. The eluted solution is evaporated to dryness in 140ଇ hot- plate to dissolve the residue 5% HNO3 for measuring ICP-MS.

Table 2. Procedure used for the ion exchange chromatography

Step Solution Concentration (M) Amount of solution (ml) Resin Cleaning DIW - 10 Conditioning HNO3 4 4 Sample loading HNO3 4 2 Washing 1 HNO3 4 4 Washing 2 HCl + Oxalic acid 5 , 0.05 6 U elution HNO3 0.1 6

The 2 mL cut of the ion exchange chromatography procedure by using multi-elemental solution is presented Fig. 1. The Recovery of Uranium was measured by ICP-MS (Nexion2000, PerkinElmer)

Fig. 1. Ion exchange chromatography elution curve of Uranium measured by ICP-MS.

3. Conclusion

Bioassay laboratory for measuring Uranium in UDGLDWLRQ ZRUNHU¶V XULQH RI .$(5, KDV EHHQ established. By using the microwave digestion method, the analytical time has been shorted and the reagent for analysis can be less consumed.

With this method, this laboratory will analyze UDGLDWLRQ ZRUNHU¶V XULQH URXWLQHO\ DQG RFFDVLRQDOO\ For further work, the ICP-MS setting for optimal measurement will be proceeded to get the reliable data from human urine.

REFERENCES

[1] Nuclear Safety And Security Commission, ³Guidelines for internal dosimetry management of radiation workers´  

>@ .HUSDV =HHY HW DO ³Uranium in Urine-QRUPDOL]DWLRQWR&UHDWLQLQH´+HDOWKSK\VLFV 86-90 (1998).

수치

Table 1. Microwave digestion program for urine samples  Step 1  Step 2  Step3  T/ଇ  R time  /min  H time /min  T/ଇ  R time /min  T/ଇ  R time /min  H time /min  120  5 3  170  5 200 7  15
Fig. 1. Ion exchange chromatography elution curve of  Uranium measured by ICP-MS.

참조

관련 문서

BIPM (International Bureau for Weights & Measures), OIML (International Organization for Legal Metrology), ILAC. (International

“With the MySQL Query Analyzer, we were able to identify and analyze problematic SQL code, and triple our database performance. More importantly, we were able to accomplish

The research tasks for achieving this purpose are: a) the experimental group's self concept will be significantly improved compared with the control group,

This study takes about 2-3 days to analyze, which is much shorter than the existing fuming nitric acid method, so it is easy to analyze a large amount of samples and will be

In this study, therefore, the method for measuring residual stresses using ESPI technique that is one of the laser applied measurement technique excellent in the view

In practice, the  equalization basin  volume will be larger 

In practice, the  equalization basin  volume will be larger  than the theoretical  value for several 

In this context, this study compared and analyzed a variety of soil decontamination methods used to evaluate the wide-area soil contaminated with Cs: