CURRENT HIGH2. VOLTAGE HIGH

문서에서 User Manual OPX-93D (페이지 42-48)

3-10. KEY LOCK

4. CURRENT HIGH2. VOLTAGE HIGH

Measuring Position

To obtain accurate result in measuring response time, peak to peak Voltage,

Load Regulation, please resistor below diagram 4-3. when load resistor is connect with B, measuring instrument must be connected to A(Binding post neck).

< Diagram 4-3 >

Front Panel Terminal Connections(Side view)

4-5. Calibration Using Front Panel

Explanation about sequence of manual calibration using Front panel CALIBRATE Key.

CALIBRATION KEY Structure

CAL 1 VOLTAGE LOW

POWER CAL

MENU

Voltage CALIBRATION Operation

▌Connect measuring instrument before calibration.

>Connect power supply (+) output terminal to DVM input terminal (+), power supply (-) output terminal to DVM input terminal (-).

▌Power ON while holding CAL Key.

Holding the key

▌Power Switch ON

▌When self-testing mode is done, it is ok to stop pressing CAL Key when

"1 CAL-VOLT LOW" is displayed Release 1.CAL VOLT LOW is displayed.

▌Stop pressing CAL Key when "1.CAL-VOLT LOW" is displayed.

LCD Display Description V-LOW 0.014V Voltage value can be different.

▌Wait for voltage shown on DVM Meter to be settled.

Release

Wait for 5 minutes

▌Use cursor key and encoder switch to input voltage value to DVM Meter.

For example, if it is 15.19mV, input as below.

LCD Display Description V-LOW 0.015V

▌Press CAL Key to save the value.

▌ y

LCD Display Description ADC DATA 026CHReadback value will be displayed in HEX for 500ms LCD Display Description 1.CAL-VOLT LOW Return to previous CAL page.

▌Use cursor key to select "2.CAL-VOLT HIGH" and press CAL Key.

LCD Display Description V-HIGH 9.251V Voltage can be different.

▌Wait for voltage shown on DVM Meter to be settled.

Wait for 5 mins

▌Use encoder switch to input voltage value shown in DVM Meter to power supply.

For example, if it is 9.2973V, input as below.

LCD Display Description V-HIGH 9.297V

▌When it is done, press CAL Key to save.

LCD Display Description ADC DATA F100HReadback value will be displayed in HEX for 500ms LCD Display Description 2.CAL-VOLT HIGH Return to previous CAL page.

▌Voltage Calibration is done, will be applied from next power supply operation.

Current CALIBRATION Operation

▌Connect measuring instrument before Calibration.

>Connect Power supply (+) terminal to electronic load input terminal (+), and connect power supply (-) to input terminal (-) current monitoring resistor, and opposite side lead to power supply (-) output terminal.

>Connect DVM input terminal (+) to Electronic load Current Monitoring resistor, input terminal (-) to opposite side of Current Monitoring resistor.

▌P P S it h ON hil i CAL K

▌Press Power Switch ON while pressing CAL Key.

▌Power Switch ON

▌When self-testing mode is done, it is ok to release CAL Key when "1.CAL-VOLT LOW" is displayed.

Holding Key

Release

▌Use cursor key to select "3.CAL-CURR LOW" and press CAL Key.

LCD Display Description A-LOW 0.0092A Current can be different.

▌Wait for voltage shown on DVM Meter to be settled.

Wait for 5 mins

▌If it is stabilized, calculate current value.

If used resist is 0.01Ω and measured voltage value is 0.0123mV, the current value will be 0.00123A.

▌Use cursor key and encoder switch to input measured current to power supply.

LCD Display Description A LOW 0 0012A LCD Display Description A-LOW 0.0012A

▌When it is done, press CAL Key to save.

LCD Display Description ADC DATA 026CHReadback value will be displayed in HEX for 500ms LCD Display Description 3.CAL-CURR LOW Return to previous CAL page.

▌Use cursor key to select "4.CAL-CURR HIGH" and press CAL Key.

LCD Display Description A-HIGH 3.2778A Current can be different.

▌Wait for voltage shown on DVM Meter to be settled.

Wait for 5 minutes

▌If it is stabilized, calculate current value.

If used resist is 0.01Ω and measured voltage value is 32.791mV, current value will be 3.2791A.

▌Use cursor key and encoder switch to input measured current value.

A HIGH 3 2 91A LCD Display Description A-HIGH 3.2791A

▌When it is done, press CAL Key to save.

LCD Display Description ADC DATA F500HReadback value will be displayed in HEX for 500ms LCD Display Description 4.CAL-CURR HIGH Return to previous CAL page.

▌Current Calibration is done and it will be applied from the next operation of power supply.

4-6. Calibration Using REMOTE INTERFACE

Explanation about calibration using Remote Interface.

Other commands cannot be used while Remote Calibration.

Connecting Measuring Instrument

▌Before calibration, connect like diagram 4-2.

▌Set up Interface in each instrument

▌Set-up Interface in each instrument.

▌Activate warming-up to all required instruments including power supply.

Remote Calibration Command Sequence

▌Read reference "6-6. Calibration Command" for SCPI Commands.

▌Please proceed with below sequence, otherwise there might be error occurs which stops Remote Calibration.

▌If an error occurs Remote Calibration should be done from the first

▌If an error occurs, Remote Calibration should be done from the first.

Voltage CALIBRATION Operation

▌Switch OFF the Electronic Load.

▌Send Voltage Minimum Calibration command to power supply.

Sending Command CAL:VOLT MIN Sending Command

▌After 5 minutes, measure DVM Voltage.

▌Send measured voltage value to power supply voltage value.

For example, if measured value is 0.1234, send as below.

Sending Command CAL:VOLT 0.1234 It is OK when digits are Over.

CAL:VOLT MIN

▌Send Voltage Maximum Calibration command to power supply.

Sending Command

▌After 5 minutes, measure DVM Voltage.

▌Send measured voltage value to power supply voltage value.

For example, if measure value is 9.123, send as below.

CAL:VOLT MAX

For example, if measure value is 9.123, send as below.

Sending Command CAL:VOLT 9.123 It is OK when digits are Over.

Current CALIBRATION Operation

▌Send Switch ON command to electronic load, and set-up current mode.

▌Send current Minimum Calibration command to power supply.

Sending Command

▌El t i L d h ld b CC M d d t th l

CAL:CURR MIN

▌Electronic Load should be CC Mode, and set-up more than power supply maximum current. If it is not more than maximum current of power supply, it can be switch to CV Mode which can bring wrong measured value, therefore, please set-up more than the maximum current.

▌Wait for 5 minutes and measure DVM current.

▌Send power supply's current value from prepared current measuring program.

For example, If measured value is 0.1234, send as below.

Sending Command It is OK when digits are Over.

▌Send Voltage Maximum Calibration command to power supply.

Sending Command

▌Wait for 5 minutes, and measure DVM Voltage connected to current monitoring resistor.

CAL:CURR MAX CAL:CURR 0.1234

▌Send measured value to power supply voltage value.

For example, if measure value is 3.1234, send as below.

Sending Command CAL:CURR 3.1234 It is OK when digits are Over.

문서에서 User Manual OPX-93D (페이지 42-48)