과산화수소 함량에 따른 시료의 부식 거동 비교

문서에서 저작자표시 (페이지 71-97)

10-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 0

2 4

6 2000ppm F + 1 % albumin

2000ppm F + 1 % albumin + 0.1 % H2O2

2000ppm F + 1 % albumin + 1 % H2O2 2000ppm F + 1 % albumin + 10 % H2O2

P ot en tia l ( V

Ag/AgCl

)

Current density (A/cm

2

)

(a)

10-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 0

2 4

6 2000ppm F + 1 % albumin

2000ppm F + 1 % albumin + 0.1 % H2O2 2000ppm F + 1 % albumin + 1 % H2O2 2000ppm F + 1 % albumin + 10 % H2O2

(b)

P ot en tia l ( V

Ag/AgCl

)

Current density (A/cm

2

)

10-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 0

2 4

(c) 6 2000ppm F + 1 % albumin

2000ppm F + 1 % albumin + 0.1 % H2O2

2000ppm F + 1 % albumin + 1 % H2O2 2000ppm F + 1 % albumin + 10 % H2O2

P ot en tia l ( V

Ag/AgCl

)

Current density (A/cm

2

)

10-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 0

2 4

(d) 6 2000ppm F + 1 % albumin

2000ppm F + 1 % albumin + 0.1 % H2O2 2000ppm F + 1 % albumin + 1 % H2O2 2000ppm F + 1 % albumin + 10 % H2O2

P ot en tia l ( V

Ag/AgCl

)

Current density (A/cm

2

)

Figure 4-24. Potential polarization curves of (c) XZ plane and (d) XY plane of AM Ti-6Al-4V alloy in the artificial saliva solution with 2000 ppm NaF + 1 wt.%

albumin + 0 ~ 10 % H2O2

0 0.1 1 10 -0.6

-0.4 -0.2 0.0

H2O2 concentration (%) (a)

SM Ti-6Al-4V YZ plane XZ plane XY plane

Ecorr (VAg/AgCl)

0 0.1 1 10

1E-5

H2O2 concentration (%) (b)

SM Ti-6Al-4V YZ plane XZ plane XY plane

icorr (A/cm2 )

10-5 10-4

(c)

SM Ti-6Al-4V YZ plane XZ plane XY plane

i1V (A/cm2 )

0 0.1 1 10 1E-5

1E-4

(d)

SM Ti-6Al-4V YZ plane XZ plane XY plane

H2O2 concentration (%) i1.6V (A/cm2 )

0 0.1 1 10

1E-5 1E-4

(e)

SM Ti-6Al-4V YZ plane XZ plane XY plane

i2.5V (A/cm2 )

H2O2 concentration (%)

Figure 4-25. (a) Ecorr(VAg/AgCl), (b) icorr(A/cm2), (c) i1V(A/cm2), (d) i1.6V(A/cm2), (e) i2.5V(A/cm2) of Ti-Al-4V alloys in the artificial saliva solution with 2000

ppm NaF + 1 wt.% albumin + 0 ~ 10 % H2O2.

Table 4.11 Ecorr(VAg/AgCl) of Ti-6Al-4V alloys in the artificial saliva solution with 2000 ppm NaF + 1 wt.% albumin + 0 ~ 10 % H2O2

Table 4.12 icorr(A/cm2) of Ti-6Al-4V alloys in the artificial saliva solution with 2000 ppm NaF + 1 wt.% albumin + 0 ~ 10 % H2O2

Solution SM Ti-6Al-4V alloy

AM Ti-6Al-4V alloy

YZ plane XZ plane XY plane AFS + 2000 ppm NaF

+ 1 wt.% albumin -0.47 -0.44 -0.34 -0.36 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 0.1 % H2O2

-0.09 -0.19 -0.12 -0.17

AFS + 2000 ppm NaF + 1 wt.% albumin

+ 1 % H2O2

-0.18 -0.22 -0.19 -0.06

AFS + 2000 ppm NaF + 1 wt.% albumin

+ 10 % H2O2

-0.12 -0.13 -0.11 -0.11

Solution SM Ti-6Al-4V alloy

AM Ti-6Al-4V alloy

YZ plane XZ plane XY plane AFS + 2000 ppm NaF

+ 1 wt.% albumin 1.02×10-5 2.43×10-6 4.44×10-6 7.2×10-6 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 0.1 % H2O2

2.90×10-6 2.87×10-6 5.49×10-6 3.93×10-6 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 1 % H2O2

4.82×10-6 4.40×10-6 5.78×10-6 7.89×10-6 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 10 % H2O2

2.82×10-5 2.61×10-5 5.89×10-5 2.64×10-5

Table 4.13 i1V(A/cm2) of Ti-6Al-4V alloys in the artificial saliva solution with 2000 ppm NaF + 1 wt.% albumin + 0 ~ 10 % H2O2

Table 4.14 i1.6V(A/cm2) of Ti-6Al-4V alloys in the artificial saliva solution with 2000 ppm NaF + 1 wt.% albumin + 0 ~ 10 % H2O2

Solution SM Ti-6Al-4V alloy

AM Ti-6Al-4V alloy

YZ plane XZ plane XY plane AFS + 2000 ppm NaF

+ 1 wt.% albumin 4.65×10-6 3.52×10-6 4.24×10-6 3.87×10-6 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 0.1 % H2O2

3.12×10-6 4.30×10-6 3.52×10-6 4.77×10-6 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 1 % H2O2

1.29×10-5 1.45×10-5 1.29×10-5 7.75×10-6 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 10 % H2O2

7.54×10-5 9.60×10-5 9.23×10-5 5.56×10-5

Solution SM Ti-6Al-4V alloy

AM Ti-6Al-4V alloy

YZ plane XZ plane XY plane AFS + 2000 ppm NaF

+ 1 wt.% albumin 7.12×10-6 1.03×10-5 7.86×10-6 8.49×10-6 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 0.1 % H2O2

1.00×10-5 8.75×10-6 8.75×10-6 1.00×10-5 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 1 % H2O2

3.15×10-5 4.27×10-5 5.66×10-5 3.25×10-5 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 10 % H2O2

2.01×10-4 2.77×10-4 3.00×10-4 2.11×10-4

Table 4.15 i2.5V(A/cm2) of Ti-6Al-4V alloys in the artificial saliva solution with 2000 ppm NaF + 1 wt.% albumin + 0 ~ 10 % H2O2

Solution SM Ti-6Al-4V alloy

AM Ti-6Al-4V alloy

YZ plane XZ plane XY plane AFS + 2000 ppm NaF

+ 1 wt.% albumin 7.1×10-6 7.23×10-6 6.73×10-6 7.05×10-6 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 0.1 % H2O2

2.67×10-5 2.52×10-5 2.37×10-5 2.67×10-5 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 1 % H2O2

1.01×10-4 9.40×10-5 1.00E-04 6.03×10-5 AFS + 2000 ppm NaF

+ 1 wt.% albumin + 10 % H2O2

4.30×10-4 4.30×10-4 5.60×10-4 3.52×10-4

(a)

(b)

Figure 4-26. Surface of (a) SM Ti-6Al-4V and (b) YZ plane after polarization test in the Artificial Saliva solution with 2000 ppm NaF + 1 wt.% albumin + 1 %

제 3 절 표면 성분 분석

1. XPS 분석

동전위 분극시험 후 표면의 화학적 상태를 분석하기 위하여 XPS 분석을 실시하였다.

시료는 -1.5 Vocp ~ 2.5 VAg/AgCl 까지 전위를 인가하며 분극시험을 실시한 뒤 즉시 꺼 내어 세척 및 건조한 시편의 표면을 분석하였으며, 적층가공 합금의 노출면에 따른 부 식저항성의 큰 차이가 없어 노출면이 YZ 면인 시료를 적층가공 Ti-6Al-4V 합금(이하 YZ plane)의 대표 군으로 하였다. 분석된 화학종은 Ti2p, O1s, F1s, Al2p, C1s 등으로 하였다.

XPS 분석의 C1s peak 결과(Figure 4-27,28.), 인공타액 내 아무것도 첨가하지 않거 나 불소 이온을 첨가하였을 경우 탄소와 탄소, 탄소와 산소의 결합량의 큰 변화는 관 찰되지 않았다. 하지만 알부민이 첨가되었을 경우 탄소 peak가 크게 증가한 것을 볼 수 있는데, 이는 단백질 요소인 알부민이 C, N, O 의 결합으로 이루어져 있어 타이타 늄 합금 표면에 생긴 탄소와 질소화합물로 예상된다.

인공타액 내 아무것도 첨가하지 않은 용액의 O1s 에서 SM Ti-6Al-4V 합금은 Al2O3 화 합물이, YZ plane 합금에서는 TiO2 결합이 더 많은 것을 확인하였다.(Figure 4-33,34.) 이는 타이타늄 뿐만 아니라 알루미늄 또한 산소친화력이 높기 때문에 분석 시료의 표면에 화합물 형태의 국부적인 양상으로 인한 편차로 판단된다. 그리고 인공 타액 내 불소이온이 첨가 되었을 때 O1s의 peak 차이는 없었다. 단, 알부민이 함유되 었을 경우 O1s peak가 현저히 줄어들었음을 알 수 있었는데 이는 인공타액 내 알부민 이 첨가되었을 경우 C1s peak 와 연관되어 설명할 수 있다. 타이타늄 표면의 산소는 탄소와 질소 등의 결합으로 O1s peak 가 낮아졌다고 판단된다. 그리고 과산화수소가 첨가되었을 경우, O1s peak 의 큰 차이는 없었다.

F1s 확인 결과(Figure 4-35,36.), 인공타액 내 불소 이온 첨가 시 SM Ti-6Al-4V 합 금과 YZ plane 합금 모두 NaTiF 및 TiF (684.85 eV), AlF (SM Ti64 - 687.24ev /

Ti-6Al-4V 합금에서 더 많이 진행된 것으로 판단된다. 하지만,SM Ti-6Al-4V 합금의 AlF3 at%는 0.2%였고, YZ plane 합금은 0.55%로 알루미늄과 불소 이온과의 화합물이 YZ plane 합금에서 더 많이 형성된 것을 알 수 있었다.

Al2p 확인 결과(Figure 4-31,32.), SM Ti-6Al-4V 합금 및 YZ plane 합금 모두 Al2p peak 가 차지하는 함량은 매우 소량이었으며, 보통 Al 또는 Al2O3 로 존재하였다. 하 지만, 인공타액 내 불소이온이 첨가된 조건에서 SM Ti-6Al-4V 합금의 경우 Al2p peak 로부터 표면에 Al(73.32 eV)이 있는 것을 알 수 있었으나, YZ plane에서 AlF3/AlF6

(75.27 eV)와 같은 알루미늄과 불소 화합물이 생성된 것을 알 수 있었다.

마지막으로 Ti2p를 확인하였을 때(Figure 4-29,30.), 인공타액에 불소 이온 첨가 시 SM Ti-6Al-4V 합금 및 YZ plane 의 표면에 Na2TiF6 (464.2 eV) 와 같은 복합화합물이 형성된 것을 알 수 있었다. 이러한 복합화합물은 부동태 피막을 불안정하게 만들어 Figure 4-23. 에 관찰된 바와 같이 부동태 전류밀도의 상승을 일으키는 요인으로 판단 된다. 또한, SM Ti-6Al-4V 합금의 경우 TiO2 가 많이 존재하며, YZ plane은 TiO, Ti2O3

가 우세함을 알 수 있었다. 그리고 인공타액 내 불소 이온과 알부민이 함께 첨가되었 을 경우 Ti2p 가 차지하는 비율이 감소하였는데, 이는 표면에 알부민이 흡착되며 막을 형성하며 Ti와 다른 원소들의 결합이 상대적으로 낮게 측정된 것으로 판단된다. 인공 타액 내 불소 이온 및 알부민과 과산화수소가 함께 첨가되었을 경우에도 SM Ti-6Al-4V 합금과 YZ plane 모두 Ti2p peak 가 현저히 줄어든 것을 확인하였다. 이는 시료 표면 에 탄소와 탄소/산소 결합 등이 주를 이루며, 상대적으로 낮은 함량의 Ti 결합이 검출 되어졌다고 판단된다. 또한, 인공타액 내 아무것도 첨가하지 않거나 불소 또는 불소/

알부민을 첨가한 용액에서는 TiO2가 Al2O3 보다 우세한 것을 알 수 있었다. 하지만 과 산화수소를 첨가하였을 시 시료 표면에 Al2O3 가 TiO2 보다 더 많은 함량을 가지고 있 었는데, 이는 기존 문헌들에 따르면 과산화수소의 첨가로 인해 용액 내 Ti, Al, V 이 용출되는데[37], 이 때 Ti의 용출이 더 많이 일어나 최종적으로 Al2O3가 더 우세한 양 상을 띤 것으로 판단된다.

SM Ti-6Al-4V YZ plane

Figure 4-27. C1s xps spectra of the surface of Ti-6Al-4V alloys after polarization test up to 2.5 VAg/AgCl in AFS ((a),(b)) and AFS + 2000 ppm NaF

SM Ti-6Al-4V YZ plane

Figure 4-28. C1s xps spectra of the surface of Ti-6Al-4V alloys after polarization test up to 2.5 VAg/AgCl in AFS + 2000 ppm NaF + 1 wt.% albumin

((a),(b)) and AFS + 2000 ppm NaF + 1 wt.% albumin + 1 % H2O2 ((c),(d))

SM Ti-6Al-4V YZ plane

Figure 4-29. Ti2p xps spectra of the surface of Ti-6Al-4V alloys after polarization test up to 2.5 VAg/AgCl in AFS ((a),(b)) and AFS + 2000 ppm NaF

((c),(d))

SM Ti-6Al-4V YZ plane

Figure 4-30. Ti2p xps spectra of the surface of Ti-6Al-4V alloys after polarization test up to 2.5 VAg/AgCl in AFS + 2000 ppm NaF + 1 wt.% albumin

((a),(b)) and AFS + 2000 ppm NaF + 1 wt.% albumin + 1 % H2O2 ((c),(d))

SM Ti-6Al-4V YZ plane

Figure 4-31. Al2p xps spectra of the surface of Ti-6Al-4V alloys after polarization test up to 2.5 VAg/AgCl in AFS ((a),(b)) and AFS + 2000 ppm NaF

((c),(d))

SM Ti-6Al-4V YZ plane

Figure 4-32. Al2p xps spectra of the surface of Ti-6Al-4V alloys after polarization test up to 2.5 VAg/AgCl in AFS + 2000 ppm F + 1 wt.% albumin ((a),(b)) and AFS + 2000 ppm NaF + 1 wt.% albumin + 1 % H2O2 ((c),(d))

SM Ti-6Al-4V YZ plane

Figure 4-33. O1s xps spectra of the surface of Ti-6Al-4V alloys after polarization test up to 2.5 VAg/AgCl in AFS ((a),(b)) and AFS + 2000 ppm NaF

((c),(d))

SM Ti-6Al-4V YZ plane

Figure 4-34. O1s xps spectra of the surface of Ti-6Al-4V alloys after polarization test up to 2.5 VAg/AgCl in AFS + 2000 ppm NaF + 1 wt.% albumin

((a),(b)) and AFS + 2000 ppm NaF + 1 wt.% albumin + 1 % H2O2 ((c),(d))

SM Ti-6Al-4V YZ plane

Figure 4-35. F1s xps spectra of the surface of Ti-6Al-4V alloys after polarization test up to 2.5 VAg/AgCl in AFS ((a),(b)) and AFS + 2000 ppm NaF

((c),(d))

SM Ti-6Al-4V YZ plane

Figure 4-36. F1s xps spectra of the surface of Ti-6Al-4V alloys after polarization test up to 2.5 VAg/AgCl in AFS + 2000 ppm NaF + 1 wt.% albumin

((a),(b)) and AFS + 2000 ppm NaF + 1 wt.% albumin + 1 % H2O2 ((c),(d))

제 5장 결 론

본 연구에서는 기존의 절삭가공 Ti-6Al-4V 합금과 노출면이 YZ, XZ, XY 면인 적층가 공 Ti-6Al-4V 합금을 인공타액 분위기에서 불소 이온, 알부민, 과산화수소의 농도에 따른 부식거동을 비교하였으며, 다음과 같은 결론을 얻었다.

1. DMT 기술로 적층제조된 Ti-6Al-4V 합금의 노출면에 따른 조직학적 차이는 있었으 나 인공타액 분위기에서 불소 이온, 알부민, 과산화수소의 첨가량에 따른 시료간의 내 식성은 유사한 수준이었다.

2. 인공타액 내 불소 이온의 농도가 증가함에 따라 시료의 부동태 전류밀도가 높아지 며, 노출면에 상관없이 적층가공 Ti-6Al-4V 합금이 절삭가공 Ti-6Al-4V 합금 대비 내 식성이 떨어지는 것을 확인하였다. 또한, 인공타액 내 고농도(2000 ppm 이상)의 불소 이온이 첨가되었을 시, 적층가공 및 절삭가공 Ti-6Al-4V 합금의 내식성은 더 큰 차이 를 보였다.

3. 2000 ppm 수준의 불소 이온이 함유된 인공타액 내 알부민 첨가 시 시료의 부식속 도를 상승시키거나 부식전위를 하강시켜 시료의 내식성에 영향을 주었다. 하지만, 타 이타늄 합금 표면에 유기화합물의 막을 형성시켜 Na2TiF6와 같은 불소화합물의 생성을 억제하는 역할을 하기도 하였다.

4. 염증상태 시 활성산소종(ROS)으로 분류되는 과산화수소의 분비는 절삭가공 및 적 층가공 Ti-6Al-4V 합금의 부식을 가속화시키는 요인으로 작용하였다.

참고 문헌

[1] Che-Haron, C. H., and A. Jawaid. "The effect of machining on surface integrity of titanium alloy Ti–6% Al–4% V." Journal of materials processing technology 166.2 (2005): 188-192.

[2] Nakagawa, Masaharu, Shigeki Matsuya, and Koich Udoh. "Corrosion behavior of pure titanium and titanium alloys in fluoride-containing solutions." Dental materials journal 20.4 (2001): 305-314.

[3] Noguti, Juliana, et al. "The role of fluoride on the process of titanium corrosion in oral cavity." Biometals 25.5 (2012): 859-862.

[4] Takemoto, Shinji, et al. "Corrosion behavior and surface characterization of titanium in solution containing fluoride and albumin." Biomaterials 26.8 (2005):

829-837.

[5] Ide, Katsuhisa, et al. "The influence of albumin on corrosion resistance of titanium in fluoride solution." Dental materials journal 22.3 (2003): 359-370.

[6] Tengvall, Pentti, et al. "Interaction between hydrogen peroxide and titanium: a possible role in the biocompatibility of titanium." Biomaterials 10.2 (1989):

118-120.

[7] Mittal, Manish, et al. "Reactive oxygen species in inflammation and tissue injury." Antioxidants & redox signaling 20.7 (2014): 1126-1167.

[8] Housley, K. L., 2007: Ch. 1 A New Element, Black Sand The History of Titanium, 1st Ed., p. 1, Metal Management Aerospace, Inc., Hartford, USA.

[9] Hunter, M. A. "METALLIC TITANIUM." Journal of the American Chemical Society 32.3 (1910): 330-336.

[10] Kroll, Wilhelm. "The production of ductile titanium." Transactions of the Electrochemical Society 78.1 (1940): 35.

[11] Kroll, W. J. "How commercial titanium and zirconium were born." Journal of the franklin institute 260.3 (1955): 169-192.

[12] Veiga, Celestino, J. P. Davim, and A. J. R. Loureiro. "Properties and

[13] Hobo, Sumiya. Osseointegration and occlusal rehabilitation. Quintessence Publishing (IL), 1989.

[14] Rubin, Leonard R. "Biomaterials in reconstructive surgery." Biomaterials in reconstructive surgery. 1983. 1017-1017.

[15] Albrektsson, T. "The response of bone to titanium implants." CRC Crit.

Rev. Biocompat. 1 (1985): 53-84.

[16] Zieliński, Andrzej, and Sylwia Sobieszczyk. "Corrosion of titanium biomaterials, mechanisms, effects and modelisation." (2008): 1-22.

[17] M. Pourbaix, Atlas of Electrochemical Equilibria in Aqueous solutions, NACE, 1974.

[18] J.Y.Lee, J.An, C.K.Chua, Fundamentals and applications of 3D printing for novel materials, Applied Materials Today, 7(2017), 120-133

[19] Ziaee, Mohsen, and Nathan B. Crane. "Binder jetting: A review of process, materials, and methods." Additive Manufacturing 28 (2019): 781-801.

[20] Sun, S., Milan Brandt, and M. J. L. A. M. Easton. "Powder bed fusion processes: An overview." Laser Additive Manufacturing (2017): 55-77.

[21] Gibson, Ian, David Rosen, and Brent Stucker. "Directed energy deposition processes." Additive manufacturing technologies. Springer, New York, NY, 2015.

245-268.

[22] Gibson, Ian, David W. Rosen, and Brent Stucker. "Sheet lamination processes." Additive Manufacturing Technologies. Springer, Boston, MA, 2010.

223-252.

[23] Lalik, Stanisław, Janusz Cebulski, and Rafał Michalik. "Corrosion resistance of titanium in water solution of hydrochloric acid." Archives of Materials Science and Engineering 28.6 (2007): 349-352.

[27] Ningshen, S., et al. "Oxygen ion irradiation effect on corrosion behavior of titanium in nitric acid medium." Journal of Nuclear Materials 408.1 (2011):

1-6.

[28] Craig, Bruce D., and David S. Anderson, eds. Handbook of corrosion data.

ASM international, 1994.

[29] Andijani, Ismaeel N., Shahreer Ahmad, and Anees U. Malik. "Corrosion behavior of titanium metal in the presence of inhibited sulfuric acid at 50 C."

Desalination 129.1 (2000): 45-51.

[30] Bäßler, Ralph, Christina Schmidt, and Helena Alves. "Recent Experiences on Suitability of Titanium and High Alloyed Materials in Acids at 200 0C."

CORROSION 2015. OnePetro, 2015.

[31] Krýsa, J., R. Mraz, and I. Roušar. "Corrosion rate of titanium in H2SO4."

Materials chemistry and physics 48.1 (1997): 64-67.

[32] Council on Dental Materials, Instruments, and Equipment. "American Dental Association status report on the occurrence of galvanic corrosion in the mouth and its potential effects." The Journal of the American Dental Association 115.5 (1987): 783-787.

[33] Goehlich, V., and M. Marek. "Corrosion behavior of PdCu and PdCo alloys in synthetic saliva." Dental Materials 6.2 (1990): 103-110.

[34] Quirynen, Marc, and C. M. L. Bollen. "The influence of surface roughness and surface‐free energy on supra‐and subgingival plaque formation in man: A review of the literature." Journal of clinical periodontology 22.1 (1995): 1-14.

[35] Bollenl, Curd ML, Paul Lambrechts, and Marc Quirynen. "Comparison of surface roughness of oral hard materials to the threshold surface roughness for bacterial plaque retention: a review of the literature." Dental materials 13.4 (1997): 258-269.

[36] Bollen, Curd ML, et al. "The influence of abutment surface roughness on plaque accumulation and peri‐implant mucositis." Clinical oral implants research 7.3 (1996): 201-211.

simulated body fluid at 25°C and 37°C. Corrosion Science, 51(10), 2473 –24 82 . Doi:10.1016/ j.corsci.2009.06.035

[38] Hsu R. W.-W., Yang C.-C., Huang C.-A. & Chen Y.-S. (2004). Investigation on the corrosion behavior of Ti-6Al-4V implant alloy by electrochemical techniques. Materials Chemistry and Physics, 86(2–3), 269–278. Doi:10.1016/

j.matchemphys.2004.02.025

[39] Sivakumar B., Kumar S. & Sankara Narayanan T.S. (2011). Fretting corrosion behaviour of Ti-6Al-4V alloy in artificial saliva containing varying concentrations of fluoride ions. Wear, 270(3–4), 317–324.

Doi:10.1016/j.wear.2010.09.008.

[40] Biesiekierski A., Ping D.H., Yamabe-Mitarai Y. & Wen C. (2014).

Impact of ruthenium on microstructure and corro- sion behavior of β-type Ti-Nb-Ru alloys for biomedical ap- plications. Materials and Design, 59, 303–

309. Doi:10.1016/ j.matdes.2014.02.058.

[41] Schutz, Ronald W. "Corrosion of titanium and titanium alloys;(ASMI Handbook Committee." Metals Handbook, Corrosion 13 (1987): 669-706.

[42] Meyer, Jean-Marc, Lucien Reclaru, and Michel Ecuvillon. "Measurement and evaluation of localized corrosion phenomena between titanium implant and dental alloys." Proceedings of the Symposium on Compatibility of Biomedical Implants, San Francisco. 1994.

[43] Huang, Her-Hsiung. "Effect of fluoride and albumin concentration on the corrosion behavior of Ti–6Al–4V alloy." Biomaterials 24.2 (2003): 275-282.

[44] Tengvall, Pentti, et al. "Interaction between hydrogen peroxide and titanium:

a possible role in the biocompatibility of titanium." Biomaterials 10.2 (1989):

118-120.

문서에서 저작자표시 (페이지 71-97)