Operation Manual

Transmitter M300 FLOW

Transmitter M300 FLOW 52 121 319

(2)
(3)

Operation Manual

Transmitter M300 FLOW

(4)

Content

1 Introduction ___________________________________________________________________________________________ 7 2 Safety instructions ______________________________________________________________________________________ 7 2.1 Definition of equipment and documentation symbols and designations ________________________________________ 7 2.2 Correct disposal of the unit _________________________________________________________________________ 8 3 Unit overview __________________________________________________________________________________________ 9 3.1 Overview 1/DIN __________________________________________________________________________________ 9 3.2 Overview 1/2DIN _________________________________________________________________________________ 10 3.3 Control/Navigation Keys ___________________________________________________________________________ 11 3.3.1 Menu Structure ___________________________________________________________________________ 11 3.3.2 Navigation keys __________________________________________________________________________ 12 3.3.2.1 Navigating the menu tree _________________________________________________________ 12 3.3.2.2 Escape _______________________________________________________________________ 12 3.3.2.3 Enter _________________________________________________________________________ 12 3.3.2. Menu _________________________________________________________________________ 12 3.3.2.5 Calibration mode _______________________________________________________________ 12 3.3.2.6 Info mode _____________________________________________________________________ 12 3.3.3 Navigation of data entry fields _______________________________________________________________ 12 3.3. Entry of data values, selection of data entry options ______________________________________________ 12 3.3.5 Navigation with ↑ in Display ________________________________________________________________ 13 3.3.6 ”Save changes” dialog _____________________________________________________________________ 13 3.3.7 Security Passwords _______________________________________________________________________ 13 3. Display ________________________________________________________________________________________ 13 4 Installation instruction __________________________________________________________________________________ 14

.1 Unpacking and inspection of equipment _______________________________________________________________ 1

.1.1 Panel cutout dimensional information – 1/DIN models ___________________________________________ 1

.1.2 Installation procedure – 1/DIN models _______________________________________________________ 15

.1.3 Panel cutout dimensional information – 1/2DIN models ___________________________________________ 16

.1. Installation procedure – 1/2DIN models _______________________________________________________ 17

.2 Connection of power supply ________________________________________________________________________ 18

.2.1 1/DIN housing (panel mount) ______________________________________________________________ 18

.2.2 1/2DIN housing (wall mount) _______________________________________________________________ 19

.3 Connector PIN definition ___________________________________________________________________________ 19

.3.1 TB1 and TB2 for 1/2DIN and 1/DIN versions __________________________________________________ 19

.3.2 TB3 and TB for 1/2DIN and 1/DIN versions __________________________________________________ 20

. Connection of Sensor _____________________________________________________________________________ 21

.5 Flow Sensor Input Wiring Kit ________________________________________________________________________ 21

.6 Kit Contents _____________________________________________________________________________________ 21

.7 Flow sensor wiring for Compatible Sensors ____________________________________________________________ 21

.7.1 Wiring for ”HIGH” type flow sensors __________________________________________________________ 22

.7.2 Wiring for ”LOW” type flow sensors ___________________________________________________________ 25

.7.3 Wiring for ”TYPE 2” flow sensors _____________________________________________________________ 26 5 Placing in/out of service ________________________________________________________________________________ 27 5.1 Placing transmitter in service _______________________________________________________________________ 27 5.2 Placing transmitter out of service ____________________________________________________________________ 27 6 Quick Setup __________________________________________________________________________________________ 28

(5)

8 Configuration _________________________________________________________________________________________ 35 8.1 Enter Configuration Mode __________________________________________________________________________ 35 8.2 Measurement Setup ______________________________________________________________________________ 36 8.2.1 Set Averaging ____________________________________________________________________________ 37 8.2.2 Set Pipe ID ______________________________________________________________________________ 37 8.2.3 Reset Total Flow __________________________________________________________________________ 38 8.2. External Total Reset _______________________________________________________________________ 38 8.3 Analog Outputs __________________________________________________________________________________ 39 8. Set Points ______________________________________________________________________________________ 0 8.5 Alarm _________________________________________________________________________________________ 2 8.5.1 Alarm Setup _____________________________________________________________________________ 2 8.6 Display ________________________________________________________________________________________ 3 8.6.1 Measurement ____________________________________________________________________________ 3 8.6.2 Resolution ______________________________________________________________________________ 3 8.6.3 Backlight _______________________________________________________________________________ 

8.6. Name __________________________________________________________________________________ 

8.7 Hold Outputs ____________________________________________________________________________________ 5 9 System ______________________________________________________________________________________________ 46 9.1 Set Language ___________________________________________________________________________________ 6 9.2 USB ___________________________________________________________________________________________ 6 9.3 Passwords _____________________________________________________________________________________ 7 9.3.1 Changing Passwords ______________________________________________________________________ 7 9.3.2 Configuring Menu Access for Operator _________________________________________________________ 7 9. Set/Clear Lockout ________________________________________________________________________________ 8 9.5 Reset __________________________________________________________________________________________ 8 9.5.1 Reset System ____________________________________________________________________________ 8 9.5.2 Reset Meter Calibration ____________________________________________________________________ 8 9.5.3 Reset Analog Calibration ___________________________________________________________________ 9 10 PID Setup ____________________________________________________________________________________________ 50 10.1 Enter PID Setup __________________________________________________________________________________ 51 10.2 PID Display Auto-Manual __________________________________________________________________________ 51 10.3 Tune Parameters _________________________________________________________________________________ 51 10.3.1 PID Assignment & Tuning __________________________________________________________________ 51 10.3.2 Setpoint & Deadband ______________________________________________________________________ 51 10.3.3 Proportional Limits ________________________________________________________________________ 52 10.3. Corner Points ____________________________________________________________________________ 52 10. Mode __________________________________________________________________________________________ 53 10..1 PID Mode _______________________________________________________________________________ 53 10..2 PID Control Tuning ________________________________________________________________________ 5

10..2.1 Tr Tuning ______________________________________________________________________ 5

11 Service ______________________________________________________________________________________________ 55 11.1 Diagnostics _____________________________________________________________________________________ 55 11.1.1 Model/Software Revision ___________________________________________________________________ 55 11.1.2 Digital Input _____________________________________________________________________________ 56 11.1.3 Display ________________________________________________________________________________ 56 11.1. Keypad ________________________________________________________________________________ 56 11.1.5 Memory ________________________________________________________________________________ 56 11.1.6 Set Relays ______________________________________________________________________________ 57 11.1.7 Read Relays _____________________________________________________________________________ 57 11.1.8 Set Analog Outputs _______________________________________________________________________ 57 11.1.9 Read Analog Outputs ______________________________________________________________________ 58 11.2 Calibrate _______________________________________________________________________________________ 58 11.2.1 Calibrate Meter ___________________________________________________________________________ 58 11.2.2 Calibrate Analog _________________________________________________________________________ 59 11.2.3 Calibrate Unlock __________________________________________________________________________ 60 11.3 Tech Service ____________________________________________________________________________________ 60 12 Info _________________________________________________________________________________________________ 61 12.1 Messages ______________________________________________________________________________________ 61 12.2 Model Software/Revision ___________________________________________________________________________ 61 13 Maintenance __________________________________________________________________________________________ 62

(6)

14 Trouble Shooting ______________________________________________________________________________________ 63 1.1 Changing the Fuse _______________________________________________________________________________ 63 15 Accessories and Spare Parts _____________________________________________________________________________ 63 16 Specifications _________________________________________________________________________________________ 64 16.1 General Specifications ____________________________________________________________________________ 6

16.2 Electrical Specifications for 1/2DIN and 1/DIN Versions __________________________________________________ 6

16.3 Mechanical Specifications for 1/DIN Versions __________________________________________________________ 65 16. Mechanical Specifications for 1/2DIN Versions __________________________________________________________ 65 16.5 Environmental Specifications for 1/2DIN and 1/DIN Versions ______________________________________________ 65 17 Default Tables _________________________________________________________________________________________ 66 18 Warranty _____________________________________________________________________________________________ 68 19 UL Statement _________________________________________________________________________________________ 68

(7)

1 Introduction

Statement of Intended Use – The M300 Flow transmitter is a single- or four- channel online process instrument for measuring conductivity or resistivity of fluids. It will interface with a variety of different Mettler-Toledo sensors, which connect to the transmitter using cables of varied lengths.

A large four line backlit Liquid Crystal Display conveys measuring data and setup information.

The menu structure allows the operator to modify all operational parameters by using keys on the front panel. A menu-lockout feature, with password protection, is available to prevent the unauthorized use of the meter. The single channel M300 transmitter can be configured to use its two analog and/or four relay outputs for process control. The four channel model uses four analog and/or six relays.

The M300 transmitter is equipped with a USB communication interface. This interface provides real-time data output and complete instrument configuration capabilities for central monitoring via Personal Computer (PC).

2 Safety instructions

This manual includes safety information with the following designations and formats.

2.1 Definition of equipment and documentation symbols and designations

WARNING: POTENTIAL FOR PERSONAL INJURY.

CAUTION: possible instrument damage or malfunction.

C

NOTE: Important operating information.

On the transmitter or in this manual text indicates: Caution and/or other possible hazard including risk of electric shock (refer to accompanying documents)

(8)

The following is a list of general safety instructions and warnings. Failure to adhere to these instructions can result in damage to the equipment and/or personal injury to the operator.

– The M300 Transmitter should be installed and operated only by personnel familiar with the transmitter and who are qualified for such work.

– The M300 Transmitter must only be operated under the specified operating conditions (see section 16).

– Repair of the M300 Transmitter must be performed by authorized, trained personnel only.

– With the exception of routine maintenance, cleaning procedures or fuse replacement, as described in this manual, the M300 Transmitter must not be tampered with or altered in any manner.

– Mettler-Toledo accepts no responsibility for damage caused by unauthorized modifications to the transmitter.

– Follow all warnings, cautions, and instructions indicated on and supplied with this product.

– Install equipment as specified in this instruction manual. Follow appropriate local and national codes.

– Protective covers must be in place at all times during normal operation.

– If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.

WARNINGS:

Installation of cable connections and servicing of this product require access to shock hazard voltage levels.

Main power and relay contacts wired to separate power source must be disconnected before servicing.

Switch or circuit breaker shall be in close proximity to the equipment and within easy reach of the OPERATOR; it shall be marked as the disconnecting device for the equipment.

Main power must employ a switch or circuit breaker as the disconnecting device for the equipment.

Electrical installation must be in accordance with the National Electrical Code and/or any other applicable national or local codes.

C

RELAY CONTROL ACTION: the M300 transmitter relays will always de-energize on loss of power, equivalent to normal state, regardless of relay state setting for powered operation.

Configure any control system using these relays with fail-safe logic accordingly.

C

PROCESS UPSETS: Because process and safety conditions may depend on consistent operation of this transmitter, provide appropriate means to maintain operation during sensor cleaning, replacement or sensor or instrument calibration.

2.2 Correct disposal of the unit

(9)

3 Unit overview

M300 models are available in both a 1/DIN and 1/2DIN case size. The 1/DIN is a panel-mount only design and the 1/2DIN models provides an integral P65 housing for wall-, or pipe-mount.

3.1 Overview 1/4DIN

4

5 6

7 8

4.01 [102]

4.01 [102]

Enter Info

METTLER TOLEDO

ESC

Menu Cal

M300

1

2 3

1 – Hard Polycarbonate case

2 – Five Tactile-Feedback Navigation Keys 3 – Four-line LCD Display

 – Power Supply Terminals 5 – USB Interface Port 6 – Relay Output Terminals

7 – Analog Output/Digital Input Terminals

8 – Sensor Input Terminals (TB  available on dual-channel units only)

(10)

3.2 Overview 1/2DIN

5.90 [150]

5.90 [150]

Enter Info Cal

Menu

M300

1

2 3

ESC

4

5

6

7 8

1 – Hard Polycarbonate case

2 – Five Tactile-Feedback Navigation Keys 3 – Four-line LCD Display

 – Power Supply Terminals 5 – USB Interface Port 6 – Relay Output Terminals

7 – Analog Output/Digital Input Terminals

8 – Sensor Input Terminals (TB  available on dual-channel units only)

(11)

3.3 Control/Navigation Keys

3.3.1 Menu Structure

Below is the structure of the M300 menu tree:

Measurement Mode M300

Cal

Menu Info

PID A/M Set Language

Measurement

Mode Passwords

Analog Outputs

Set/Clear Lockout Calibrate

Set Points

Reset Tech Service

Alarm

Display

Hold Outputs

Diagnostics

Tune Parameters USB

Measurement Set Averaging Set Pipe ID Reset Total Flow External Total Reset

Measurement Resolution

Backlight Name

Model/Software Revision Digital Input

Display Keypad Memory Set Relays Read Relays Set Analog Outputs

Read Analog Outputs

Messages

Service Model/Software

Revision System

Configure

Quick Setup PID Setup

(12)

3.3.2 Navigation keys

Menu Cal Info Enter

ESC

3.3.2.1 Navigating the menu tree

Enter the desired main Menu branch with the ,  or  keys. Use the  and  keys to navigate through the selected Menu branch.

C

NOTE: In order to back up one menu page, without escaping to the measurement mode, move the cursor under the UP Arrow character at the bottom right of the display screen and press [Enter].

3.3.2.2 Escape

Press the  and  key simultaneously (escape) to return to the Measurement mode.

3.3.2.3 Enter

Use the

key to confirm action or selections.

3.3.2.4 Menu

Press the  key to access the main Menu.

3.3.2.5 Calibration mode Press the  key to enter Calibration Mode.

3.3.2.6 Info mode

Press the  key to enter Info Mode

3.3.3 Navigation of data entry fields

Use the  key to navigate forward or the  key to navigate backwards within the changeable data entry fields of the display.

(13)

3.3.5 Navigation with ↑ in Display

If a ↑ is displayed on the bottom right hand corner of the display, you can use the  or the  key to navigate to it. If you click [ENTER] you will navigate backwards through the menu (go back one screen). This can be a very useful option to move back up the menu tree without having to exit into the measuring mode and re-enter the menu.

3.3.6 ”Save changes” dialog

Three options are possible for the ”Save changes” dialog: Yes & Exit (Save changes and exit to measuring mode), ”Yes & ↑” (Save changes and go back one screen) and ”No & Exit”

(Don’t save changes and exit to measuring mode). The ”Yes & ↑” option is very useful if you want to continue configuring without having to re-enter the menu.

3.3.7 Security Passwords

The M300 transmitter allows a security lock-out of various menus. If the security lock-out feature of the transmitter has been enabled, a security password must be entered to allow access to the menu. See section 9.3 for more information.

3.4 Display

C

NOTE: In the event of an alarm or other error condition the M300 Transmitter will display a flashing in the upper right corner of the display. This symbol will remain until the condition that caused it has been cleared.

C

NOTE: During calibrations, clean, Digital In with Analog Output/Relay/USB in Hold state, a flashing H will appear in the upper left corner of the display. This symbol will remain for 20 seconds until after the calibration or clean is completed .This symbol will also disappear when Digital In is deactivated.

(14)

4 Installation instruction

4.1 Unpacking and inspection of equipment

Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions.

Do not discard the box.

If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present.

If items are missing, notify Mettler-Toledo immediately

4.1.1 Panel cutout dimensional information – 1/4DIN models

1/DIN Model transmitters are designed for panel-mount installation only. Each transmitter is supplied with mounting hardware to provide fast and simple installation to a flat panel or flat enclosure door. To insure a good seal and maintain IP integrity of installation, the panel or door must be flat and have a smooth finish. Hardware consists of:

Two – Snap-on Mounting brackets One – Mounting gasket seal

Transmitter dimensions and mounting are shown in figures below.

3.62 +0.02

–0

[92 +0.5]

–0

3.62 +0.02

–0

[92 +0.5]

–0

PANEL CUT-OUT

(15)

4.1.2 Installation procedure – 1/4DIN models

– Make cutout in panel (see dimensions cutout drawing).

– Be sure surface surrounding cutout is clean, smooth and free of burrs.

– Slide face gasket (supplied with transmitter) around transmitter from the back of the unit.

– Place transmitter into cutout hole. Be sure there are no gaps between the transmitter and panel surface.

– Place the two mounting brackets on either side of the transmitter as shown – While holding transmitter firmly into the cutout hole, push the mounting brackets

toward the backside of panel

– Once secure, use a screwdriver to tighten the brackets against the panel – Face gasket will compress between transmitter and panel

CAUTION: Do not over tighten brackets

4.01 [102]

4.33 [110]

3.55 [90]

4.01 [102]

3.55 [90]

4.01 [102]

4.96 [126]

1.77 [45]

3.54 [90]

4.33 [110]

MAX. 0.118 [3]

FLAT GASKET MOUNTING SCREW 2 PLACES

MOUNTING BRACKET 2 PLACES

Power Fuse

USB 1 2 3 4 5 6 7 1 2 3 4 5 6 7

1 2 3 4 5 6 7 8 9

1 2 3 4 5 6 7 8 9

4.92 [125]

4.01 [102]

4.33 [110]

3.55 [90]

4.01 [102] 3.55

[90]

4.01 [102]

4.96 [126]

1.77 [45]

3.54 [90]

4.33 [110]

MAX. 0.118 [3]

FLAT GASKET MOUNTING SCREW 2 PLACES

MOUNTING BRACKET 2 PLACES

Power Fuse

USB 1 2 3 4 5 6 7 1 2 3 4 5 6 7

1 2 3 4 5 6 7 8 9

1 2 3 4 5 6 7 8 9

4.92 [125]

(16)

4.1.3 Panel cutout dimensional information – 1/2DIN models

1/2DIN Model transmitters are designed with an integral rear cover for stand-alone wall mount installation.

The unit may also be wall mounted using the integral rear cover. See installation instructions in Section .1..

Below are cut-out dimensions required by the 1/2DIN models when mounted within a flat panel or on a flat enclosure door. This surface must be flat and smooth. Textured or rough surfaces are not recommended and may limit the effectiveness of the gasket seal provided.

5.39

+0.02 –0.00

[137

+0.5 –0.0

]

5.39

+0.02 –0.00

[137

+0.5 –0.0

]

PANEL CUT-OUT

Optional hardware accessories are available that allow for panel- or pipe-mount.

Refer to Section 15 for ordering information.

(17)

4.1.4 Installation procedure – 1/2DIN models

For Wall Mount:

– Remove rear cover from front housing.

– Start by unscrewing the four screws located on the face of the transmitter, in each corner.

This allows the front cover to swing away from the rear housing.

– Remove the hinge-pin by squeezing the pin from each end.

This allows the front housing to be removed from the rear housing – Drill out wall-mount breakthroughs in the rear housing.

– Mount rear housing to wall using appropriate mounting hardware for wall surface. Be sure it is level and securely fastened and the installation adheres to any and all clearance dimensions required for transmitter service and maintenance.

– Insert two black protective covers (supplied with the M300 transmitter) over the fixing hardware and into the space on the inside back cover, as shown in the drawing below.

This is necessary to maintain unit integrity.

– Replace the front housing to the rear housing. The unit is ready to be wired.

For Pipe Mount:

– Use only manufacturer-supplied components for pipe-mounting the M300 transmitter and install per the supplied instructions. See section 15 for ordering information.

5.90 [150]

5.90 [150]

0.256 [6.5]

3.30 [84]

1.65 [42]

4.56 [116]

1.13 [28.7]

1.15 [29.2]

3.54 [90]

2.95 [75]

0.236 [6]

1.38 [35]

3.15 [80]

5.36 [136]

5.36 [136]

CABLE GLAND OR CONDUIT BREAKTHROUGHS ø1/2" (ø 21.5mm)

PIPE MOUNTING BREAKTHROUGHS (4 PLACES)

WALL MOUNTING BREAKTHROUGHS (2 PLACES)

CABLE GLANDS (3 PLACES)

ESC

Enter Info Cal Menu

METTLER TOLEDO M300

(18)

4.2 Connection of power supply

All connections to the transmitter are made on the rear panel of all models.

Be sure power to all wires is turned off before proceeding with the installation. High voltage may be present on the input power wires and relay wires.

A two-terminal connector on the rear panel of all M300 models is provided for power

connection. All M300 models are designed to operate from a 20–30 VDC or a 100 to 20 VAC power source. Refer to specifications for power requirements and ratings and size power wiring accordingly.

The terminal block for power connections is labeled ”Power” on the rear panel of the transmitter.

One terminal is labeled –N for the Neutral wire and the other +L for the Line (or Load) wire.

There is no earth ground terminal on the transmitter. For this reason the internal power wiring within the transmitter is double insulated and the product label designates this using the

h symbol.

4.2.1 1/4DIN housing (panel mount)

1

2

1 Connection of power supply 2 Terminal for sensor

(19)

4.2.2 1/2DIN housing (wall mount)

1 2

1 Connection of power supply 2 Terminal for sensor

4.3 Connector PIN definition

4.3.1 TB1 and TB2 for 1/2DIN and 1/4DIN versions

Power connections are labeled – N for Neutral and +L for Line, for 100 to 240 VAC or 20 – 30 VDC.

14DIN 12DIN

* Dual channel only TB1A for 14DIN

1 NO2 2 COM2 3 NC2 4 NO6*

5 COM6*

6 NO4 7 COM4

TB1B for 14DIN 1 NO1 2 COM1 3 NC1 4 NC5*

5 COM5*

6 NO3 7 COM3 TB2 for 14DIN

1 AO1+

2 AO1–/AO2–

3 AO2+

4 AO3+*

5 AO3–/AO4–*

6 AO4+*

7 DI1+

8 DI1–/DI2–*

9 DI2+*

TB2 for 12DIN 1 AO1+

2 AO1–/AO2–

3 AO2+

4 AO3+*

5 AO3–/AO4–*

6 AO4+*

7 DI1+

8 DI1–/DI2–*

9 DI2+*

1 NO1 2 COM1 3 NC1 4 NO2 5 COM2 6 NC2 7 COM5*

8 NC5*

9 COM6*

10 NO6*

11 NO3 12 COM3 13 NO4 14 COM4 TB4

TB2 TB3

TB1A

1 7 TB1B

1 7

9

1 9

1 9

1

TB2 TB3 TB4

TB11 14

1 9 1 9 1 9

TB1 for 12DIN

NO = normally open (contact open if un-actuated).

NC = normally closed (contact closed if un-actuated).

(20)

4.3.2 TB3 and TB4 for 1/2DIN and 1/4DIN versions

TB3 and TB are used for sensor inputs.

TB3

Pin no. Transmitter Function

TB3 TB*

1 – Not used

2 GND Ground

3 BJ* DJ* +10 VDC

 Bin* Din* Flow Pulse Input

5 +5V + 5 VDC

6 GND Ground

7 AJ CJ* + 10 VDC

8 AIn CIn* Flow Pulse Input

9 +5V + 5 VDC

* Four channel models only

AJ and AIN refer to connections for channel A.

(21)

4.4 Connection of Sensor

The M300 FLOW Transmitter is designed to operate with various types of sensors. These sensors require different wiring configurations. Listed below are instructions for wiring the various types of sensors offered by Mettler-Toledo Thornton for use with this transmitter. Please consult the factory for assistance if attempting to wire sensors not offered by Mettler-Toledo Thornton as some sensors may not be compatible.

4.5 Flow Sensor Input Wiring Kit

This kit contains components that may be needed at input terminals to condition sensor signals.

Refer to the following sections or to the instruction manual for wiring details.

4.6 Kit Contents

This kit contains the following items:

− x Wire nuts

− x 10K ohm resistors for use with Burket 8020 and 8030 type sensors, and GF Signet 2500-series sensors.

− x 1K ohm resistors for use with Data Industrial 200-series and Fluidyne insertion type sensors.

− x 0.33uF, 50 V capacitors for use with Berket 8020 and 8030 type sensors, Data Industrial 200-series and 000-series sensors, GF Signet 2500-series sensors, Sanitary Turbine-Type sensors, Fluidyne insertion type sensors and Racine Federated (Formerly Asahi/

America) vortex-style sensors.

4.7 Flow sensor wiring for Compatible Sensors

The following sections provide wiring information to properly connect various compatible flow sensors to the M300 FLOW Transmitter. When using the Configuration menu of the transmitter to setup the flow sensor, the first prompt asks to select the TYPE of flow sensor being connected.

There are three choices as follows:

High: All flow sensors described in Section .5.1

Low: P515 Signet flow sensors only, described in section .5.2 Type 2: Asahi flow sensors, described in Section .5.3

(22)

4.7.1 Wiring for ”HIGH” type flow sensors

The following wiring information is used when connecting (Burkert 8020 and 8030 type) inline Hall effect 5VDC, flow sensors. Thornton models 33901 thru 33935.

THORNTON 33901-33935

Burkert 8020, 8030 Hall-Effect Sensor

Type: high M300 transmitter

L–, shield Pulse output L+

10K 0.33uF

1TB 3 2 GND 3 45 +5V 6 GND 7 A J 8 A IN 9 +5V

Extension cable not provided. Use 2-conductor twisted pair with shield, 22 AWG (Belden 851 or equivalent), 1,000 ft (305 m) maximum length.

The following wiring information is used when connecting Badger (formerly Data Industrial 200-Series) forward-swept paddlewheel type flow sensors.

Thornton models 33142 thru 33145 and 33159 thru 33162 and 33273.

THORNTON 33142-33145, 33159-33162, 33273

Badger

(formerly Data Industrial 200 Series) Flow Sensors

Type: high

M300 transmitter black

shield

red 1K

0.33uF

1TB 3 2 GND 34 5 +5V 6 GND 7 A J 8 A IN 9 +5V

Extension cable provided with sensor. Use 2-conductor twisted pair with shield 20AWG (Belden 9320 or equivalent) to extend length to 2000 ft (610 m) max.

(23)

The following wiring information is used when connecting Badger (formerly Data Industrial

000-Series) forward-swept paddlewheel type flow sensors.

Thornton models 33174 thru 33177 and 33171 and 33172.

THORNTON 33174-33177, 33171, 33172

Badger

(formerly Data Industrial 4000 Series) Flow Sensors

Type: high M300 transmitter

black shield red clear

0.33uF

1TB 3 2 GND 3 45 +5V 6 GND 7 A J 8 A IN 9 +5V

20 ft (6.1 m) extension cable provided with sensor. Use 3-conductor with shield, 20 AWG (Belden 936 or equivalent) to extend length to 2000 ft (610 m) maximum.

The following wiring information is used when connecting (GF Signet 2500-Series) Hall Effect paddlewheel type flow sensors. Thornton models 33282, 33285, 33287, 33298 thru 33305.

THORNTON 33282, 33285, 33287, 33298-33305

Signet 2507, 2536, 2540 Hall-Effect

Paddlewheel Sensor, Type: high M300 transmitter

shield red black

10K 0.33uF

1TB 3 2 GND 34 5 +5V 6 GND 7 A J 8 A IN 9 +5V

25 ft (7.6 m) extension cable provided with sensor. Use 2-conductor with shield, 22 AWG (Belden 851 or equivalent) to extend length to 1000 ft (305 m) maximum.

(24)

The following wiring information is used when connecting Sanitary Turbine type flow sensors.

Thornton models 33336 thru 33377 (Hoffer) and 33441 thru 33450 (Sponsler).

THORNTON 33336-333377 (Hoffer)

Hoffer Turbine Sensors

Type: high M300 transmitter

black shield red white

0.33uF

1TB 3 2 GND 3 45 +5V 6 GND 7 A J 8 A IN 9 +5V

20 ft (6.1 m) extension cable provided with sensor. Use 3-conductor with shield, 20 AWG (Belden 936 or equivalent) to extend length to 3000 ft (915 m) maximum.

THORNTON 33441-33450 (Sponsler)

Sponsler Turbine Sensor

Type: high M300 transmitter

black shield red orange

0.33uF

1TB 3 2 GND 34 5 +5V 6 GND 7 A J 8 A IN 9 +5V

20 ft (6.1 m) extension cable provided with sensor. Use 3-conductor with shield, 20 AWG (Belden 936 or equivalent) to extend length to 3000 ft (915 m) maximum.

(25)

The following wiring information is used when connecting Spirax Sarco/Emco flow (formerly Fluidyne) insertion type flow sensors. Thornton models 33358 thru 33375.

THORNTON 33358-33375

Spirax Sarco/Emco flow (formerly Fluidyne) Insertion Vortex Sensor

Type: high

M300 transmitter (–)

GND

(+) 1K

0.33uF

1TB 3 2 GND 3 45 +5V 6 GND 7 A J 8 A IN 9 +5V

Extension cable not provided. Use 2-conductor twisted pair with shield, 20 AWG (Belden 9320 or equivalent), 2000 ft (610 m) maximum length.

4.7.2 Wiring for ”LOW” type flow sensors

The following wiring information is used when connecting (GF Signet 515) type flow sensors.

Thornton models 33189, 33193, 33195, 33196, and 33229.

THORNTON 33189, 33193, 33195, 33196, 33229

Signet 515 Sensor

Type: low M300 transmitter

black shield red

1TB 3 2 GND 34 5 +5V 6 GND 7 A J 8 A IN 9 +5V

Extension cable not provided. Use 2-conductor twisted pair with shield, 22 AWG (Belden 851 or equivalent, 200 ft (61 m) maximum length.

(26)

4.7.3 Wiring for ”TYPE 2” flow sensors

The following wiring information is used when connecting Racine Federated (formerly Asahi/

America) vortex flow sensors. Thornton models 33308 to 33335.

THORNTON 33308-33335

Racine Federated (formerly Asahi/America)

Vortex Sensor

Type: 2 M300 transmitter

(–) Ground (+) Power Signal

0.33uF

1TB 3 2 GND 3 45 +5V 6 GND 7 A J 8 A IN 9 +5V

Extension cable not provided. Use 3-conductor with shield, 20 AWG (Belden 936 or equivalent), 1000 ft (305 m) maximum length.

(27)

5 Placing in/out of service

5.1 Placing transmitter in service

After connecting the transmitter to power supply circuit, it will be active as soon as the circuit is powered.

5.2 Placing transmitter out of service

First disconnect the unit from the main power source then disconnect all remaining electrical connections. Remove the unit from the wall/panel. Use the installation instruction in this manual as reference for dis-assembling mounting hardware.

(28)

6 Quick Setup

(PATH: Menu/Quick Setup)

Quick Setup allows limited configuration of the most common functions of the M300 Flow transmitter. Detailed information for each function can be found in the individual sections of the manual.

6.1 Enter Quick Setup mode

Select Quick Setup and press the [ENTER] key. Enter the security code if necessary (see section 9.3)

C

Note: Refer to section 3.3 for information on menu navigation.

While in Measurement mode press the [MENU] key to bring up the Menu selection. Select Quick Setup and press the [ENTER] key.

Convention:

1st line on display => a 2nd line on display => b 3rd line on display => c

th line on display => d

Only lines a and b on single channel models or lines a and c on dual channel models can be configured in Quick Setup. Go to the Configuration Menu to configure remaining lines.

6.2 Flow Sensor Type Selection

Refer to Section .5 for Flow sensor type information. Select the desired flow sensor type.

When configuring four channel transmitters, also select the channel A, B, C or D to be con- figured. Press [ENTER]

6.3 Calibration Constant Entry

(29)

6.4 Measurement Selection

Select the desired display line (a or c) of the single channel transmitter to configure the values to be displayed and whether this value will have an Analog Output.

Convention (single channel models):

1st line on display => a 3rd line on display => c Example:

By selecting a and GPM as units, the flow rate value will be displayed on the 1st line.

By selecting c and Gals as units, the total flow value will be displayed on the 3rd line of the display which has high resolution.

Selecting None means the display will be blank for the line selected.

Four channel transmitters, configure the channel followed by the measurement units.

Convention (four channel models only):

1st line on display => Channel A 2nd line on display => Channel B 3rd line on display => Channel C

th line on display => Channel D

6.5 Analog Outputs

By selecting Analog Output Yes on the previous screen a linear –20 mA analog output Aout will be setup for the measurement when [ENTER] is pressed. Selecting No means that no analog output is setup. Aout min and Aout max are the minimum and maximum measurement values for the  and 20mA values respectively.

Analog Output defaults for Quick Setup:

Measurement a => Aout 1 Measurement c => Aout 2 Measurement A => Aout 1 Measurement B => Aout 2 Measurement C => Aout 3 Measurement D => Aout 

(30)

6.6 Set Points

After configuring the Analog Output a Set Point can be configured for that output. If No is selected and [ENTER] is pressed then the Quick Setup is done and the menus are exited without setting any Set Point.

Selecting Yes and pressing [ENTER] means a Set Point can be configured. Following Set Point Types can be selected:

Off (Set Point is Off)

High (High value has to be set) Low (Low value has to be set)

Outside (High and Low value has to be set) Between (High and Low value has to be set)

Total Flow (Available only if units of total flow are chosen. A Total Flow value has to be set) After setting the Set point value(s) a Relay (none (blank), 1, 2, 3, ) can be configured for that Set Point. The Relay delay is set to 10 seconds and the Hysteresis is set to 5%.

Press [ENTER] to bring up the Save Changes dialog.

(31)

7 Calibration

(PATH: Cal)

The calibration key  allows the user one-touch access to the calibration features of the transmitter.

7.1 Enter Calibration Mode

While in Measurement mode press the  key. Enter the security code if necessary (see section 9.) Press the  or  key to select the type of calibration desired. The options are ”Sensor”,

”Meter”, or ”Analog”.

C

NOTE: To exit Calibration mode at any time press the  and  keys simultaneously (Escape).

The transmitter returns to the Measurement mode and the old calibration values remain active.

C

NOTE: If only Calibrate Sensor function is accessible, the Calibrate Unlock function is set to ”No”.

To access all calibration functions using the one-touch Cal key, go to Service/Calibrate menu and change Unlock setting to ”Yes”. Refer to Section 11.2 for Meter and Analog calibration instructions and for instructions to unlock calibration functions for Cal key. [Analog and Meter calibration are always accessible in the Service/Calibrate menu]

C

NOTE: During calibration, the outputs will default to be held at their current values until 20 seconds after the calibration menu is exited. A flashing H appears in the upper left corner of the display while outputs are held. Refer to Section 8.7 Hold Outputs to change the hold output status.

(32)

7.2 Sensor Calibration

This feature provides the ability to perform a One-point or Two-point Sensor flow calibration and

”Edit” or ”Verify” saved calibration constants. The most common method of calibration for flow sensors is to enter the calibration constants appropriate for the sensor using the Edit function.

Some users may choose to perform an in-line calibration using a One-point or Two-point Sensor flow calibration. This requires an external reference system. When performing an in-line cali- bration on a flow sensor, results will vary depending on the methods and calibration apparatus used to perform the calibration.

Select the channel (four channel models only) and the desired calibration option. Choices are GPM, meters/hour or liters/minute (for a one-pt or two-pt flow calibration), Edit and Verify. Press [ENTER]

7.2.1 One point calibration

Select 1 point Calibration by pressing the  or  key followed by the [ENTER] key.

Enter the Value of calibration Point 1 from the external reference system and press the [ENTER]

key to start calibration. The value in the 2nd text line is the actual measured value from the sensor prior to calibration.

After the calibration, the slope (Multiplier) calibration factor M and the offset (Adder) calibration factor A are displayed.

Select Yes to save the calibration values and the successful Calibration is confirmed on the display. Select No to discard the entered calibration. The meter retains the old calibration values and returns to the Measurement mode.

(33)

7.2.2 Two point Calibration

Enter the Sensor Calibration mode as described in section 7.2. Select 2 point Calibration followed by the [ENTER] key.

Enter the Value of Point 1 from the external reference system and press [ENTER]. Change the flow rate to another value. For best results, the change in flow rate should be as large as practical. The change in flow rate may be either high to low or low to high.

Enter the Value of Point 2 from the external reference system and press [ENTER] to start the calibration.

After the calibration, the slope (Multiplier) calibration factor M and the offset (Adder) calibration factor A are displayed.

Select Yes to save the net calibration values and the successful Calibration is confirmed on the display. Select No to discard the entered calibration. The meter retains the old calibration values and returns to the Measurement mode.

(34)

7.3 Edit

The Edit function is the most commonly used calibration method for flow sensors.

Enter Calibration mode as described in section 7.1 and select the channel (four channel models only) and select Edit.

Press [ENTER] to display all calibration constants for the sensor. The calibration constants can be changed in this menu. If the sensor type previously selected was either High or Low, the M and A values will be displayed. If Type 2 was selected a table of K and F values will be displayed.

Press [ENTER] until prompted to save calibration values

Select Yes to save the new calibration values and the successful Calibration is confirmed on the display.

7.4 Verify

Enter Calibration mode as described in section 7.1 and select the channel (four channel models only) and Verify.

The measurement value and the frequency (Hz) are shown. The meter calibration factors are used when calculating the measurement value.

Press [ENTER] to return to the Measurement mode.

(35)

8 Configuration

(PATH: Menu/Configure)

Configure

Measurements Set Averaging Set Pipe ID Reset Total Flow External Total Reset

Measurement Resolution

Backlight Name Analog

Outputs Set Points Alarm/Clean Display Hold Outputs Measurement

C

NOTE: Screen shots represent typical single channel displays. Displays for four channel models may vary.

8.1 Enter Configuration Mode

While in Measurement mode press the  key. Press the  or  key to navigate to the Configure – Menu. Select the Configure – Menu and enter the security code ”xxxxx” if necessary (see section 9.). Press the [ENTER] key to confirm the code.

C

NOTE: to exit Configuration mode at any time press the  and  key simultaneously (escape).

The transmitter returns to the Measurement mode and the old settings remain active.

(36)

8.2 Measurement Setup

(PATH: Menu/Configure/Measurement)

Press [ENTER] to select Measurement Menu. The following sub menus can now be selected:

Measurements, Set Averaging, Set Pipe ID, Reset Total Flow and External Total Reset.

Press [ENTER] to select Measurements.

Select the type of sensor(s) wired to the transmitter and press [ENTER]. The options are High, Type 2 or Low. See Section .5 for sensor types.

The  lines of the display can now be configured with a value. When configuring single channel transmitters, pressing the [ENTER] key will display the selection for lines c and d.

Convention, Single Channel:

1st line on display => a 2nd line on display => b 3rd line on display => c

th line on display => d

Four channel transmitters allow the configuration of both primary (A, B, C, D) and secondary values (a, b, c, d). Press [ENTER] to display channels B through D.

C

NOTE: Pressing [ENTER] during normal Measuring mode of four channel transmitters will toggle the display between the Primary and Secondary values.Convention,

 Channel:

1st line on display => A (a) 2nd line on display => B (b) 3rd line on display => C (c)

th line on display => D (d)

Pressing the [ENTER] key again will bring up the Save Changes dialog.

(37)

8.2.1 Set Averaging

Press [ENTER] to select this Menu. The averaging method (noise filter) for each measurement can now be selected. The options are Special (Default), None, Low, Medium and High.

None = no averaging or filtering

Low = equivalent to a 3 point moving average Medium = equivalent to a 5 point moving average High = equivalent to a 7 point moving average

Special = averaging depending on signal change (ideal for large changes in input signal).

Pressing the [ENTER] key will scroll through the remaining measurements.

Press[ENTER] to bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made.

8.2.2 Set Pipe ID

Press [ENTER] key to select this Menu. This menu is used for flow sensors where the inside diameter of the pipe where the flow sensor is installed, is needed to calculate an accurate flow velocity measurement.

Enter the inside diameter of the pipe. The value can be entered in (inches) or (cm).

Press [ENTER] to configure channels C and D of four channel transmitters.

Press [ENTER]to bring up the Save Changes dialog.

(38)

8.2.3 Reset Total Flow

Press [ENTER] key to select this Menu. This menu is used to reset the totalized flow value.

Select Reset Total Yes or No for each sensor channel. Press [ENTER] to display sensor channels C and D of four channel transmitters.

Press [ENTER] to bring up the Save Changes dialog.

8.2.4 External Total Reset

Press [ENTER] key to select this Menu. This menu is used to reset the totalized flow value using the digital input feature of the transmitter.

Select reset total (flow), Yes or No. Four channel transmitters must also select the digital input to use (either 1 or 2). Pressing [ENTER] will bring up channels C and D.

Press [ENTER] to bring up the Save Changes dialog.

(39)

8.3 Analog Outputs

(PATH: Menu/Configure/Analog Outputs)

Enter configuration mode as described in Section 8.2

Press the [ENTER] key to select this Menu, which allows configuration of the Analog Outputs Two Analog Outputs are available for single channel transmitters and  on four channel units.

Once the analog outputs have been selected, use the  and  buttons to navigate between configurable parameters. Once a parameter is selected, its setting can be selected per the following table:

Parameter Selectable Values

Aout: 1, 2, 3* or * (default is 1)

Measurement: a, b, c, d or blank (none) (default is blank) Alarm Value: 3.6 mA, 22.0 mA of Off (default is off)

* Aout 3 and  available on two-channel units only

When an Alarm Value is selected, the analog output will go to this value if any alarm condition occurs

The Aout type can be Normal, Bi-Linear, Auto-Range or Logarithmic. The range can be –20mA or 0–20mA. Normal provides linear scaling between the minimum and maximum scaling limits and is the default setting. Bi-Linear will also prompt for a scaling value for the mid-point of the signal and allows two different linear segments between the minimum and maximum scaling limits.

Enter the minimum and maximum Value of Aout.

If Auto-range was selected then Aout max1 can be configured. Aout max1 is the maximum value for the first range on Auto-Range. The maximum value for the second range on Auto- Range was set in the previous menu. If Logarithmic Range was selected, it will also prompt for the number of decades as ”Aout1 # of Decades =2”.

The value for the Hold mode can be configured to hold the last value or can be set to a Fixed value.

Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made.

(40)

8.4 Set Points

(PATH: Menu/Configuration/Set Points)

Press the [ENTER] key to select this Menu. This menu is used to configure Setpoints.

Up to  Set Points for single channel transmitters and 8 for four channel transmitters can be configured on one measurement on this screen. Types are Off, High, Low, Outside, Between and Total Flow.

Off (Set Point is Off)

High (High value must be set) Low (Low value must be set)

Outside (High and Low value must be set) Between (High and Low value must be set)

Total Flow (Available only if units of total flow are chosen. A Total Flow value must be set) Enter the desired value(s) for the Set Point and press [ENTER].

This screen provides the option to configure a setpoint to be active on an over range condition.

Select the setpoint (1 thru ) and ”Yes” or ”No”. Select the desired relay that will activate when the setpoint alarm condition is reached.

Over Range

Configure whether an over range should also be alarmed and which relay should be used. Once configured, the selected relay will be activated if a sensor over-range condition is detected on the assigned input channel.

Delay

Enter the delay time in seconds. A time delay requires the setpoint to be exceeded continuously for the specified length of time before activating the relay. If the condition disappears before the delay period is over, the relay will not be activated.

Hysteresis

Enter the hysteresis as a percentage-value. A hysteresis value requires the measurement to return within the setpoint value by a specified percentage before the relay is deactivated.

For a high setpoint, the measurement must decrease more than the indicated percentage below the setpoint value before the relay is deactivated. With a low setpoint, the measurement must rise at least this percentage above the setpoint value before the relay is deactivated. For example, with a high setpoint of 100, when this value is exceeded, the measurement must fall below 90 before the relay is deactivated.

(41)

State

Relay contacts are in normal state until the associated setpoint is exceeded, then the relay is activated and the contact states change.

Select ”Inverted” to reverse the normal operating state of the relay (i.e. Normally open contacts are in a closed state, and normally closed contacts are in an open state, until the setpoint is exceeded). ”Inverted” relay operation is functional when power is applied to the M300 trans- mitter.

Four channel transmitters also allow configuration of a Relay Hold Status of ”Last” or ”Off”.

This is the state the Relay will go to during a Hold status.

Press [ENTER] to bring up the Save Changes dialog.

(42)

8.5 Alarm

(PATH: Menu/Configuration/Alarm)

This Menu allows the configuration of an Alarm.

8.5.1 Alarm Setup

To select ”Setup Alarm”, press the  or  key so that ”Alarm” is flashing.

Using the  and  buttons, navigate to ”Use Relay #”. Using the  or  keys, select relay (1, 2, 3 or ) to be used for the Alarm and press [ENTER].

One of the following events may be alarmed:

1. Power Failure 2. Software Failure

If any of these are set to Yes then the alarm will come on and an alarm message will be recorded if:

1. there is a power failure or power cycling 2. the software watchdog performs a reset

For 1 and 2 the alarm indicator will be turned off when the alarm message is cleared. It will reappear if the power is constantly cycling or if the watchdog is repeatedly resetting the system.

Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values, selecting Yes will make the entered values the current ones

C

Note: Each Alarm Relay can be configured in either a Normal or Inverted state. In addition, a Delay for the activation can be set. For more information, refer to Section 8.5.

(43)

8.6 Display

(PATH: Menu/Configure/Display)

Enter configuration mode as described in Section 8.1.

This Menu allows for the configuration of the values to be displayed and also the configuration of the Display itself.

8.6.1 Measurement

The Display has  lines. Line 1 on top and Line  on the bottom.

Select the values (Measurement a, b, c or d) to be displayed on each line of the display.

Select the ”Error Display” mode. If this is set to ”On” when an alarm has occurred, the message

”Failure – Press Enter” will be displayed on Line  when an alarm occurs in the normal Measurement mode.

C

NOTE: Pressing [ENTER] during normal Measuring mode of four channel transmitters will toggle the display between the Primary (A–D) and Secondary (a–d) values.

Select the ”Error Display” mode. If this is set to ”On” when an alarm has occurred, the message

”Failure – Press Enter” will be displayed on Line  when an alarm occurs in the normal Measurement mode.

Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values, selecting Yes will make the entered values the current ones.

8.6.2 Resolution

This menu allows the setting of the resolution of each displayed value.

Possible settings are 1, 0.1, 0.01, 0.001 or Auto.

Pressing the [ENTER] key Will bring up the Save Changes dialog.

(44)

8.6.3 Backlight

This Menu allows setting the Backlight options of the display.

Possible settings are On, On 50% or Auto Off 50%. If Auto Off 50% is selected then the backlight will go to 50% after  minutes with no keypad activity. The backlight will automatically come back on if a key is pressed.

Pressing the [ENTER] key again will bring up the Save Changes dialog.

8.6.4 Name

This menu allows for the configuration of an alpha-numeric name which is displayed on Lines 3 and  of the Display. The default is nothing (blank).

Use the  and  keys to navigate between digits to be altered. Using the  and  keys to change the character to be displayed. Once all digits of both display channels have been entered, press [ENTER] to bring up the Save Changes dialog.

Names can also be displayed on the secondary Measurement mode screen of four channel transmitters. Pressing [ENTER] allows configuration of the secondary names.

Press [ENTER] again to bring up the Save Changes dialog.

참조

관련 문서

In the Simple mode, macrocells are configured as dedicated inputs or as dedicated, always active, combinatorial outputs.. Architecture configurations available in this mode

Additive alignment, which occurs after passing orifice channel, has been visualized, showing additive alignment that is perpendicular to the flow direction at the

In Moore machines, outputs are dependent only on the present state, the output will change synchronously if combinational logic

• Flow in the reservoir and central core of the tube: primary forces are pressure and gravity forces.. • Selective withdrawal: Colder water is withdrawn into the intake channel

In this paper, the power line characteristics as the view point of communication channel are investigated and the technologies of PLC are proposed to

In this paper, we propose a method to guarantee the system life as passive elements by using high efficiency operation and analog devices by tracking

‰ Sequential logic circuits’ outputs react to the current values on the input wires and to the past history of values on those same input wires....

In this study, we fabricated electrode tools for electrochemical machining (ECM) using three-dimensional (3D) printing, which outputs free shapes