Quantitative Analysis of Intraoperative Indocyanine Green Video Angiography in Aneurysm Surgery

Young-Je Son1, Jeong Eun Kim2, Sung Bae Park1, Sang Hyung Lee1, Young Seob Chung1, Hee-Jin Yang1

1Department of Neurosurgery, Seoul National University Boramae Hospital, Republic of Korea

2Department of Neurosurgery, Seoul National University College of Medicine, Seoul, Republic of Korea

Objective : Indocyanine green (ICG) videoangiography (VA) is being used in assessment of blood flow during cerebrovascular surgery. However, da- ta collected during ICG angiography are usually interpreted qualitatively.

In this study, quantitative analysis of ICG angiogram was attempted.

Materials and Methods : ICG VA, performed during aneurysm surgery was analyzed retrospectively. The angiogram was captured serially in reg- ular time interval. The stacked images were then fed into an image anal- ysis program, ImageJ. The selected areas of interest were as follows: pa- rent and branch vessels, and dome of aneurysm. Changes of signals of measurement points were plotted. The time to peak, washout time, and the peak intensity between areas were compared.

Results : Among the 16 cases enrolled in this study, five cases were an- terior communicating artery aneurysms, and 11 cases were middle cere- bral artery bifurcation aneurysms. There was no signal intensity of aneur- ysm dome in our series. No difference in time to peak or maximum sig- nal intensity was observed between vessels in each case. The average time to peak was 9.0 and washout time was 31.3 seconds. No significant difference in time profile was observed between anterior communicating artery aneurysms and middle cerebral artery bifurcation aneurysms.

Conclusion : Findings of this study demonstrate that quantitative analysis is possible using a personal computer and common video capture and analysis software. It can be a good adjunctive to evaluation of vascular status during aneurysm surgery. It displays time profiles of multiple points of interest over time, and is helpful in objective evaluation of changes of blood flow over time. It might be helpful in various fields of cere- brovascular surgery.

J Cerebrovasc Endovasc Neurosurg.

2013 June;15(2):76~84 Received : 10 March 2013 Revised : 17 May 2013 Accepted : 11 June 2013 Correspondence to Hee-Jin Yang Department of Neurosurgery, Boramae Hospital, Seoul National University 20, Boramae-ro 5-gil, Dognjak-gu, Seoul, 156-707, Republic of Korea

Tel : 82-2-870-2303 Fax : 82-2-870-3863 E-mail : nsyang@brm.co.kr

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non- Commercial License (http://creativecommons.org/li- censes/by-nc/3.0) which permits unrestricted non- commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Keywords Indocyanine green video angiography, Aneurysm, Quantitative analysis, ImageJ

INTRODUCTION

Since the first report on intraoperative assessment of vascular flow using indocyanine green (ICG) video an- giography (VA) by Raabe et al.,15) a number of articles describing its use in various fields of vascular surgery

have been reported. However, most reported articles de- scribed qualitative analysis, and only a few articles have described quantitative analysis.1-2)14) Most of the studies reporting on quantitative analysis used specialized hard- ware or proprietary software, so that application of the described techniques to one's own study is not easy.

(2)

Case No Age Sex Location Peak (s) Wash-out (s)

1 71 F MCA 6.0 50.0

2 53 M Acom 22.0 44.0

3 58 M MCA 6.0 36.0

4 67 F MCA 6.0 18.0

5 55 F MCA UC UC

6 58 F MCA 11.0 27.0

7 56 F MCA 8.0 19.0

8 53 M MCA UC UC

9 41 M Acom UC UC

10 49 F MCA 8.0 28.0

11 46 F MCA UC UC

12 67 F Acom 4.4 9.8

13 35 M MCA UC UC

14 62 M Acom 5.0 56.4

15 69 F Acom UC 24.0

16 32 M MCA 14.0 32.0

M= male; F= female; MCA= middle cerebral artery bifurcation aneurysm; Acom= anterior communicating artery aneurysm; s= seconds; UC=

uncheckable

Table 1. Summary of cases - Demographics, locations of aneurysms and time profiles.

formed during aneurysm surgery from January 2010 to December 2010, was conducted. ICG VA was per- formed using a commercially available surgical micro- scope (OPMI® PenteroTM, Carl Zeiss Co, Oberkochen, DE). After intravenous bolus injection of ICG (Daiichi Sankyo, Tokyo, JP), the field of interest was illumi- nated with near-infrared light. ICG, which had been stored at room temperature in a shaded area, was mixed with 10 ml of designated solvent just before injection. After confirmation of complete dissolution of ICG power in solvent, the mixed solution was in- fused through a central venous line, under the direc- tion of an operator. Real-time angiographic image was seen on a video screen and recorded for further analysis. At each time, a standard dose of 25 mg dis-

files with an interval from 0.01 to 120 seconds.

The image stack generated by capture software was fed into ImageJ (Ver. 1.42q, http://rsb.info.nih.gov /ij/, NIH, USA), using a multi-measure plug-in. The regions of interest (ROI) were selected for analysis, and the intensity changes of each ROI were then cal- culated by ImageJ as a spread sheet file. Selected ROIs were as follows: parent artery, distal branches to aneurysm, and dome of aneurysm. The ROI's were drawn in a circular shape to cover the vascular area on an averaged image. Averaged images were gen- erated by the z-stack function of ImageJ. The spread sheet was used to make a graphic plot of changes in intensity of each ROI over time.

The graphs were reviewed and factors for achieve-

(3)

A B C

D E F

Fig. 1. Comparison of graph details according to sampling interval. A, D are made in 0.1 seconds; B, E in 0.2 seconds; C, F in 0.4 seconds. The waves in the lower row (D, E, and F) are five times magnified view of a part of waves, shaded area, in the upper row (A, B, and C). It is notable that there is some loss of detail in the 0.4 second interval compared to 0.1 and 0.2 second inter- vals (F). The horizontal bar in A, B, and C is 10 seconds, in D, E, and F is 2 seconds.

ment of better results were sought: the interval of se- rial capture, condition of recorded video. In addition, the time characteristics of graphs, time to peak, and washout time were also evaluated. Time to peak was defined as the interval from uprise of a wave to high- est signal intensity of a wave, and washout time was defined as the interval from uprise of a wave to the time when rapid fall of signal intensity was changed to slow downfall. Statistical analysis was performed us- ing SPSS (version 20, IBM SPSS Statistics, Armonk, NY).

RESULTS

Demographics

A total of 16 cases with ICG VA were analyzed.

There were five cases of anterior communicating ar-

tery aneurysms and 11 cases of middle cerebral artery aneurysms (Table 1). There were seven males and nine females. The patients' ages were 54.5 ± 11.7 years (mean ± standard deviation, range 32 to 71).

Parameters for Analysis

In order to set up the best conditions for analysis, a graph was made with serial captures of 0.1, 0.2, and 0.4 second intervals. Routine evaluation found no def- inite differences in waves. However, using the magni- fied view, some loss of detail was observed in 0.4 sec- ond interval waves, compared to that of 0.1 and 0.2 second intervals (Fig. 1). Therefore, serial capture for analysis was performed using 0.2 second intervals.

The analysis was performed using a desktop per- sonal computer without any delay in image process-

(4)

D E F

Fig. 2. Example of good data quality (A, B, and C) and of poor quality (D, E, and F). A and D are photographs of operative fields, B and E are averaged images of serial captures of indocyanine green (ICG) videoangiography (VA), C and F are graphs.

There is no blurring of contours of structures in B (1: M2 inferior division, 2: M2 superior division, arrow: aneurysm) whereas E shows blurring due to panning of the field during recording (M1: middle cerebral artery main trunk proximal to bifurcation, M2S:

superior division of M2, Arrowhead: inferior division of M2, arrow: aneurysm). C (M2I: inferior division of M2, M2S: superior division of M2, An: aneurysm) shows good waveforms, compared to waves in F (M1: middle cerebral artery main trunk proximal to bifurca- tion, M2S: superior division of M2, M2I: inferior division of M2, An: aneurysm).

ing and analysis. It runs under Windows XP pro- fessional (Korean version), equipped with AMD® AthlonTM CPU (64×2 Dual 5600+), 4GB RAM, NVIDIA® GeForce 8600GTTM graphic card.

Evaluation of the quality of data by superimposing image stacks was helpful. This process was performed using the z-project function of ImageJ. The averaged image showed no blurring when the video had been recorded without panning or zooming of the oper- ative field and the graph showed good quality appro- priate for analysis, whereas data with blurring of averaged image showed poor quality of graph (Fig. 2).

Evaluation of time profile

The time profile of each graph was assessed. Time to peak and washout time were measured from each graph (Fig. 3).

A summary of the measured time profile is shown in Table 1. In some cases, the wave was not suitable for assessment of time profile. The cause of unsuitable data was change of magnification (zooming) and moving of the operation field (panning) during ICG-VA recording, which precluded obtaining a static field for immobile ROIs. It was denoted as uncheckable. No definite difference in time profile was observed between proximal and distal vessels in valid cases (Fig. 4). There was no significant signal in- tensity of aneurysm dome in our series.

The time to peak was 9.0 ± 5.4 (mean ± standard de- viation), and washout time was 31.3 ± 14.3 seconds. No significant difference in time profile was observed be- tween anterior communicating artery aneurysms and middle cerebral artery bifurcation aneurysms (Table 2).

(5)

Fig. 3. Time profile of wave. 1: time to peak, 2: washout time. Arrowhead marks the point of transition of the curve. In this case, time to peak is 24.0 seconds, washout time 44.0.

A B C

D E F

Fig. 4. A case of an anterior communicating artery aneurysm. A is a photograph of the operative field. B to F are graphs of meas- urement depicted in A. B is a graph of point 1 (A1); C, point 2 (ipsilateral A1); E, point 3 (contralateral A2); F, point 4 (aneurysm).

D is combined graphs B, C, E, and F. There is no difference in time profile between vessels, whereas no definite signal intensity at the aneurysm.

(6)

Location N* Mean Standard deviation P value**

Time to

peak (s) Acom 3 10.5 10.0 0.615

MCA 7 8.45 3.0

Washout

time (s) Acom 4 33.6 20.7 0.713

MCA 7 30.0 10.9

* Number of valid cases

** independent sample t test

Table 2. Comparison of peak and wash-out time according to aneurysm location

a normal vascular permeability. Because it is rapidly eliminated exclusively by the liver (half-life, 2.4 mi- nutes), repeated injections of ICG are possible.3)

ICG VA is based on the fluorescence of ICG. In plasma, it displays an absorption maximum at 805 nm and an emission maximum at 830 nm.3) Adverse re- actions related to its clinical use are rare, and side ef- fects, other than iodine allergy, have not been reported.3)9) A dose of 0.2 to 0.5 mg/kg is recom- mended for ICG-VA, with a maximum daily dose limit of 5 mg/kg.4)18)

Since the report on intraoperative flow assessment during cerebrovascular surgery by Raabe et al., a number of studies using ICG VA in various fields of neurosurgery, such as arteriovenous malformation (AVM), dural arteriovenous fistula (AVF), carotid en- darterectomy (CEA), and extracranial-intracranial ar- terial bypass (EIAB), in addition to aneurysms, have been reported.1-2)8)10-11)16-17) These studies reported that ICG-VA was very useful for assessment of parent ar- tery patency, detection of aneurysm dome filling, as- sessment of the completeness of the removal of the AVM nidus, localization and confirmation of complete AVF obliteration, and determining the residual steno- sis in CEA.

ysis have been reported.1-2)6)12-14) These studies re- ported on the usefulness of quantitative analysis of ICG-VA in prediction of survival of flaps and evalua- tion of change in cardiac muscle perfusion area ac- cording to progression of vascular compromise.6)12-13) Quantitative analysis has several merits. It can dif- ferentiate residual ICG signal in the dome of an aneurysm from real re-filling of ICG by evaluating changes of signal intensities over time. It also is help- ful in detection of vascular compromise according to difference in time profile between vessels.14) Difference in time profile between vessels is helpful for identi- fication of feeding artery, draining vein, and normal cortical vein in AVM resection.2)

All of the reported articles describing quantitative analysis used specialized hardware or proprietary software, preventing other researchers from perform- ing analysis on their own data. This study demon- strated the feasibility of quantitative analysis using conventional hardware and software. GomPlayer, free software popular in Korea, was used for serial cap- ture of ICG-VA. However, any software capable of se- rial capture can be used instead. Measurement of seri- al ROI was performed using ImageJ, free software distributed by the National Institutes of Health (NIH).

(7)

Any software can be used to make graphs presented in this article; for example, Microsoft Excel (Microsoft Corp., Seattle, WA).

Although ImageJ is capable of handling video file directly, its performance is often limited by various codec's used in video files. We think that it is more comfortable to use image stacks (set of serially cap- tured images) rather than using the video file as it is. There might be a limitation to the number of im- ages captures in some software; therefore, it would be better to perform serial capture using the longest interval possible. Increased number of serially cap- tured images might also require more system resources. We found that a capture interval of 0.2 seconds per frame is sufficient for preservation of de- tails in waves (Fig. 1).

The hardware specifications used for analysis in our study were of an intermediate level desktop computer purchased in 2008. Considering the rapid progress in computer technology, the analysis can be performed on almost any current desktop computer or notebook computer. Therefore, the technique described in this article is open to anyone who wishes to analyze the ICG-VA by himself.

One of the requirements for better analysis is the static field, which means no motion or change of magnification of the operative field during ICG-VA recording (Fig. 2). All of the data depicted as 'un- checkable' in Table 1 resulted from this unstatic field of view. The critical time profile, required for analy- sis, was displayed in 120 seconds (2 minutes) after the first appearance of fluorescence signal in arteries.

Therefore, maintenance of a static field for 2 to 3 mi- nutes is recommended as sufficient for acquiring an ICG-VA suitable for analysis. If the operator wants to see a higher magnification view or check the vascular patency at a different angle, it is better to wait for 2 to 3 minutes or redo the ICG-VA with a second shot of dye in case of need.

No significant difference in time profile was ob- served between proximal and distal branch vessels.

Considering the proximity of points of measurement and velocity of arterial flow, this finding is thought to be not changed in larger series. A difference in time profile between vessels, although not depicted in our series, might suggest serious compromise in blood flow in that vessel. No definite difference of time pro- files among vessels was reported in an EIAB series.1) No significant difference in time profile was observed between anterior communicating artery aneurysms and middle cerebral artery bifurcation aneurysms.

This means that there is no definite difference in he- modynamics between cerebral arteries, although this finding might need to be verified with a larger series.

Our data showed large variation in time to peak and washout time. Among the 16 cases in our series, there were six cases with poor quality data. Therefore, the actual number of cases used for analysis of time pro- file was approximately two thirds of the total cases.

We think that the high variation might be a reflection of a small number of cases and due in part to the suboptimal quality of data.

Intraoperative microvascular Doppler sonography (mDs) is another adjunctive now frequently used for quality assessment in cerebrovascular surgery. mDS is easy to apply with instant results; however, because it only evaluates a pinpoint site, comparison of differ- ences between points is difficult.12) ICG-VA is good for simultaneous evaluation of perfusion of areas with multiple points, however, it is not suitable for evalua- tion of blood flow of giant, complex, and deep-seated aneurysms.4) mDs and ICG-VA have been reported to complement one another, and combination of the two techniques assures the quality of aneurysm surgery.7)

Our data include several features. This study was conducted using conventional hardware and software for quantitative analysis. Any ICG-VA, recorded ac- cording to the guidelines described, can be used as source data for analysis, as described in the Materials and Methods section. It described the time profile of ICG-VA in aneurysm surgery, which might provide baseline data for future study with a larger series.

(8)

verified with those of larger series. Data elucidated by large series might be helpful for detection of minute derangements from normal value, and will make quantitative analysis more useful for improvement of surgical outcome of cerebrovascular surgeries. In ad- dition, application of quantitative analysis of ICG-VA might be useful in various aspects of cerebrovascular surgery other than aneurysms.

CONCLUSION

Quantitative analysis of ICG-VA is feasible using a personal computer and common video capture and analysis software. It displays time profiles of multiple points of interest over time, helps in objective evalua- tion of changes of blood flow over time, and it might be helpful in various fields of cerebrovascular surgery.

CONFLICTS OF INTEREST

The authors do not have any conflict of interest.

ACKNOWLEDGEMENTS

This study was supported by a research grant of Seoul National University Boramae Hospital.

REFERENCES

1. Awano T, Sakatani K, Yokose N, Kondo Y, Igarashi T, Hoshino T, et al. Intraoperative EC-IC bypass blood flow assessment with indocyanine green angiography in moyamoya and non-moyamoya ischemic stroke. World Neurosurg. 2010 Jun;73(6):668-74.

neurosurgical treatment of intracranial aneurysms: A review. Acta Neurochir Suppl. 2010;107:107-9.

5. de Oliveira JG, Beck J, Seifert V, Teixeira MJ, Raabe A.

Assessment of flow in perforating arteries during intra- cranial aneurysm surgery using intraoperative near-infra- red indocyanine green videoangiography. Neurosurgery.

2007 Sep;61(3 Suppl):63-72; discussion 72-3.

6. Detter C, Wipper S, Russ D, Iffland A, Burdorf L, Thein E, et al. Fluorescent cardiac imaging: a novel intra- operative method for quantitative assessment of my- ocardial perfusion during graded coronary artery stenosis.

Circulation. 2007 Aug;116(9):1007-14.

7. Fischer G, Stadie A, Oertel JM. Near-infrared indocyanine green videoangiography versus microvascular Doppler sonography in aneurysm surgery. Acta Neurochir (Wien).

2010 Sep;152(9):1519-25.

8. Hanggi D, Etminan N, Steiger HJ. The impact of mi- croscope-integrated intraoperative near-infrared in- docyanine green videoangiography on surgery of arte- riovenous malformations and dural arteriovenous fistulae. Neurosurgery. 2010 Oct;67(4):1094-103; dis- cussion 1103-4.

9. Hope-Ross M, Yannuzzi LA, Gragoudas ES, Guyer DR, Slakter JS, Sorenson JA, et al. Adverse reactions due to indocyanine green. Ophthalmology. 1994 Mar;101(3):529-33.

10. Lee CH, Jung YS, Yang HJ, Son YJ, Lee SH. An in- novative method for detecting surgical errors using in- docyanine green angiography during carotid endarter- ectomy: a preliminary investigation. Acta Neurochir (Wien). 2012 Jan;154(1):67-73.

11. Lee JC, Jang KS, Jang DK, Han YM, Park SK, Yun WS, et al. Analysis of incomplete occlusion of cerebral aneurysm by intraoperative indocyanine green videoangiography.

Korean J Cerebrovasc Surg. 2010 Sep;12(3):206-12.

12. Matsui A, Lee BT, Winer JH, Laurence RG, Frangioni JV. Quantitative assessment of perfusion and vascular compromise in perforator flaps using a near-infrared flu- orescence-guided imaging system. Plast Reconstr Surg.

2009 Aug;124(2):451-60.

13. Mothes H, Dinkelaker T, Donicke T, Friedel R, Hofmann GO, Bach O. Outcome prediction in microsurgery by quan- titative evaluation of perfusion using ICG fluorescence angiography. J Hand Surg Eur Vol. 2009 Apr;34(2):238-46.

14. Oda J, Kato Y, Chen SF, Sodhiya P, Watabe T, Imizu S, et al. Intraoperative near-infrared indocyanine green-vid- eoangiography (ICG-VA) and graphic analysis of fluo- rescence intensity in cerebral aneurysm surgery. J Clin Neurosci. 2011 Aug;18(8):1097-100.

(9)

15. Raabe A, Beck J, Gerlach R, Zimmermann M, Seifert V.

Near-infrared indocyanine green video angiography: a new method for intraoperative assessment of vascular flow. Neurosurgery. 2003 Jan;52(1):132-9; discussion 139.

16. Schuette AJ, Cawley CM, Barrow DL. Indocyanine green videoangiography in the management of dural arterio- venous fistulae. Neurosurgery. 2010 Sep;67(3):658-62; dis- cussion 662.

17. Taddei G, Tommasi CD, Ricci A, Galzio RJ. Arteriovenous malformations and intraoperative indocyanine green video- angiography: preliminary experience. Neurol India. 2011 Jan-Feb;59(1):97-100.

18. Wang S, Liu L, Zhao Y, Zhang D, Yang M, Zhao J.

Evaluation of surgical microscope-integrated intraoperative near-infrared indocyanine green videoangiography during aneurysm surgery. Neurosurg Rev. 2010 Apr;34(2):209-15.

수치

Table 1. Summary of cases - Demographics, locations of aneurysms and time profiles.
Fig. 1. Comparison of graph details according to sampling interval. A, D are made in 0.1 seconds; B, E in 0.2 seconds; C, F in 0.4  seconds
Fig. 2. Example of good data quality (A, B, and C) and of poor quality (D, E, and F). A and D are photographs of operative  fields, B and E are averaged images of serial captures of indocyanine green (ICG) videoangiography (VA), C and F are graphs
Fig. 3. Time profile of wave. 1: time to peak, 2: washout time. Arrowhead marks the point of transition of the curve
+2

참조

관련 문서

Second, there was no significant difference in the effect of participants' motivation according to age, final educational attainment, motivation to

Second, analyzing students' level of interest in the unit of Clothes and Life Style revealed that there was significant difference between genders and students in the two

Measurement of proximal contact tightness between the left first molar and second molar(dental implant) in the mandible... Clinical dental anatomy, histology,

Sixth, there was no significant correlation between lower extremity muscle thickness, static stability, and dynamic stability of excellent athletes in college

2) In between-group comparison before and after elastic band exercise, a significant difference was found in the percentage of body fat, fat-free

As for the number of participation, significant difference was found in accessibility, attractiveness and comfort of physical environment, recognition and

There was no significant difference between the groups in problems of eating habits, but the obesity group considered overeating as a problem while the normal

Conclusions : No significant corelation of the patency rate was shown in the study on the effects of local irrigation and systemic application of heparin